CB-S / CB-S-UL (E7) 06/2020
page 40/119
7.
Temperature and CO
2
set-point entry
Required access level: “User”.
Setting ranges
Control ranges
Temperature
0 °C /
32 °F
up to 50 °C /
122 °F
6 °C /
10.8 °F
above ambient up to 50 °C /
122 °F
Humidity
0 vol.-% up to 20 vol.-%
0 vol.-% up to 20 vol.-%
7.1
Temperature set-point entry
Path:
Normal display
Setpoints
Temperature
Press the
OK button
to enable the setting.
Temperature setting.
The current setting flashes. Enter the desired set-point with the
arrow
buttons
.
Confirm the entry with the
OK button
.
Temperature
With the
arrow-down button
you can continue with the CO
2
set-point entry (chap. 7.2).
With the
Back button
you can go back to the “
Setpoints
” submenu and, repeatedly pressing it, to
Nor-
mal display
.
When setting a lower temperature set point, in order to save time, we recommend cooling
down the chamber by turning it off and opening the chamber doors.
With safety controller mode “Limit”, adapt the safety controller always when you changed the
temperature set-point. Set the safety controller value by approx. 2 °C above the temperature
set-point (chap. 11.2).
Recommended setting: safety controller mode “Offset” with safety controller value 2 °C.
7.2
CO
2
set-point entry
Path:
Normal display
Setpoints
CO2
Press the
OK button
to enable the setting.
CO
2
set-point setting
The current setting flashes. Enter the desired set-point with the
arrow
buttons
.
Confirm the entry with the
OK button
.
CO2
With the
arrow-up button
you can go back to the temperature set-point entry (chap. 7.1)
With the
arrow-down button
you can now change to the special controller functions setting (chap. 9)
With the
Back button
you can go back to the “
Setpoints
” submenu and, repeatedly pressing it, to
Nor-
mal display
.
When setting a lower CO
2
set point, the CO
2
gas must be able to escape first. Open the
chamber doors for this purpose.
Summary of Contents for 9040-0189
Page 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Page 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Page 112: ...CB S CB S UL E7 06 2020 page 112 119...
Page 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...