4.5
Electrical connection
The chambers are supplied ready for connection. They come with an IEC connector plug.
Model
Power plug
Nominal v/- 10% at the
indicated power frequency
Current
type
Chamber
fuse
CB-S 170
Grounded plug
IEC 7/7
200-230 V at 50 Hz
200-230 V at 60 Hz
1N~
10 A
CB-S 170-UL
NEMA 5-20P
100-120 V at 50 Hz
100-120 V at 60 Hz
1N~
16 A
CB-S 260
Grounded plug
IEC 7/7
200-230 V at 50 Hz
200-230 V at 60 Hz
1N~
10 A
CB-S 260-UL
NEMA 5-20P
100-120 V at 50 Hz
100-120 V at 60 Hz
1N~
16 A
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the lat-
est technology. The protective conductors of the socket and plug must be compatible!
DANGER
Electrical hazard due to missing protective conductor connection.
Deadly electric shock.
Make sure that the chamber’s power plug and the power socket match and secure-
ly connect the electrical protective conductors of the chamber and the house instal-
lation.
•
Only use original connection cables from BINDER according to the above specification.
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (centrally located at the bottom of the left-hand side of the
chamber, see chap. 1.6).
NOTICE
Danger of incorrect power supply voltage due to improper connection.
Damage to the chamber.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
•
When connecting the chamber, please observe the regulations specified by the local electricity supply
company as well as the local and national regulations (VDE directives for Germany).
•
Make sure that there is sufficient current protection in accordance with the number of devices that are
to be operated. We recommend the use of a residual current circuit breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
See also electrical data (chap. 26.4).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Summary of Contents for 9040-0189
Page 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Page 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Page 112: ...CB S CB S UL E7 06 2020 page 112 119...
Page 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...