14. Chamber settings (for experienced users only)
14.1 Setting the humidity control
The humidifying system provides a maximum humidity of up to 95 % r.h. in the inner chamber, which
remains condensation-free. This performance assumes an average ambient temperature of +22 °C +/- 3
°C /
71.6 °F +/- 5.4 °F
and a working temperature in the inner chamber of 37 °C /
98.6 °F
.
Depending on the installation site and ambient temperature you can increase or decrease the humidity by
up to +/- 5 % r.h. if required. Factory setting: 0 % r.h.
If the ambient temperature deviates by more than +/- 5 °C from the values recommended by
the manufacturer, the conditions for maximum air humidity with condensation-free inner
chamber are no longer guaranteed. Contact BINDER Service for assistance.
Path:
Normal display
Settings
Various
Humidity con-
trol
Press the
OK button
to enable the setting.
Setting the humidity control
The current setting flashes. Enter the desired value with the
arrow
buttons
. Entry range: -5 % r.h. up to +5 % r.h.
Confirm the entry with the
OK button
.
Humidity control
With the
arrow-up button
you can change to the door open alarm delay setting (chap. 13.4).
With the
arrow-down button
you can now change to the door heating offset setting.
With the
Back button
you can go back to the “
Various
” submenu and, repeatedly pressing it, to
Normal
display
.
14.2 Setting the door heating offset
In unfavorable ambient conditions, condensation may occur in the door area. To avoid this, an Offset
value for the door heating can be set. Factory setting: 3.0 °C.
If the ambient temperature deviates by more than +/- 5 °C from the values recommended by
the manufacturer, the conditions for maximum air humidity with condensation-free inner
chamber are no longer guaranteed. Contact BINDER Service for assistance.
Path:
Normal display
Settings
Various
Doortemp.
offset
Press the
OK button
to enable the setting.
Setting the door temperature offset.
The current setting flashes. Enter the desired value with the
arrow
buttons
. Entry range: -5.0 °C up to +5.0 °C.
Confirm the entry with the
OK button
.
Doortemp. offset
With the
arrow-up button
you can go back to the humidity control setting.
With the
Back button
you can go back to the “
Various
” submenu and, repeatedly pressing it, to
Normal
display
.
Summary of Contents for 9040-0189
Page 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Page 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Page 112: ...CB S CB S UL E7 06 2020 page 112 119...
Page 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...