DIH 0300 B – USER GUIDE REANIBEX 300
54
8 CONFIGURATION
Figure 35: Confirmation request during the Hardware Test
.
5. Check that the results of all the tests are OK. If any errors are detected, please contact the
Authorised Technical Service.
6. To exit this menu option, press the key
.
NOTE:
It is recommended not to perform this Hardware test more than 5 consecutive times to prevent the
internal components of the equipment from undergoing excessive temperatures.
8.7.2 ACCESSORIES
TEST
This option is used to check whether the REANIBEX 300 is capable of correctly detecting the type of
patient and electrode cables that are connected to it.
When this option is accessed, a screen appears indicating the type of electrodes that are connected
(adult or paediatric), the type of patient cable (3 leads) and its status.
8.7.3 USER INTERFACE TEST
This option is used to check that all the buttons and indicators on the user interface work correctly.
When this option is accessed, a screen is displayed representing the user interface of the REANIBEX 300
with all its buttons and indicators.
This is an interactive test that requires user participation. When the various user interface buttons of the
equipment are pressed, the graphic representation on the screen lights up and a beep is heard. If the
button has an associated light indicator, the indicator lights up as long as the button is pressed.
Take the following steps to perform the User Interface Test:
1. Switch on the REANIBEX 300 in Configuration mode.
2. Use the programmable buttons to access the “TESTS” option
.
3. Access the “USER INTERFACE TEST” option. The user interface test screen will appear.
4. Press the various buttons on the user interface of the equipment to check that the graphic lights
up. If the equipment does not respond when one or more buttons are pressed, contact the
Authorised Technical Service.