DIH 0300 B – USER GUIDE REANIBEX 300
87
A.6 ELECTROMAGNETIC COMPATIBILITY GUIDE
exceeds the applicable RF level of compliance, the REANIBEX should be monitored to check that it
operates normally. If operation is not normal, additional measures should be taken, such as changing the
orientation or location of the REANIBEX 300.
RECOMMENDED DISTANCES BETWEEN MOBILE AND PORTABLE RF COMMUNICATION
EQUIPMENT AND THE REANIBEX 300
The REANIBEX 300 is intended for use in settings where RF interferences are controlled. The client or
user of the REANIBEX 300 can help to prevent electromagnetic interference by maintaining a minimum
distance between mobile and portable RF communication equipment (transmitters) and the REANIBEX
300, as recommended below, depending on the maximum output power of the communication
equipment.
Maximum
transmitter
output power
W
Separation according to transmitter frequency (m)
150 KHz to 80
MHz outside
ISM bands
d =1.2 P
150 KHz to 80
MHz within ISM
bands
d =1.2 P
80 MHz to 800
MHz
d =1.2 P
800 MHz to 2.5
GHz
d=2.3 P
0.01 0.12 0.12 0.12 0.23
0.1 0.38 0.38 0.38 0.73
1 1.2 1.2 1.2 2.3
10 3.8 3.8 3.8 7.3
100 12 12 12 23
For transmitters with a maximum output power not specified above, the recommended distance of
separation d in meters (m) can be calculated by using the equation applicable to the frequency of the
transmitter, where P is the maximum output power in watts (W) according to the manufacturer of the
transmitter.
NOTE 1
: The distance of separation for the highest frequency is applied at 80 Mhz and 800 MHz.
NOTE 2
: An additional factor of 10/3 is used in the calculation of the separation distances recommended
for the 80 MHz to 2.5 GHz frequency range. This reduces the probability that the mobile/portable
communications equipment will cause interferences if it is inadvertently placed in a patient area.
NOTE 3
: These guiding measures may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection of buildings, objects and people.