67
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5% of the value shown
Measurement inaccuracy
Max. permissible standard deviation according to clinical testing:
Systolic 8 mmHg/diastolic 8 mmHg
Memory
2x 60 memory spaces (With 30-second measurement. Deviation possi-
ble in the event of a longer measurement duration.)
Dimensions
Main unit: L 120 mm x W 144 mm x H 53 mm
ECG stick: D 22 mm x H 120 mm
Weight
Main unit: Approximately 490 g (without batteries, with cuff)
ECG stick: Approximately 40 g
Cuff size
22 to 42 cm
Permissible operating cond10 °C to +40 °C, 10 to 85 % relative humidity (non-condensing)
Permissible storage and transport
conditions
-20 °C to +55 °C, 10 to 90 % relative humidity,
800 – 1050 hPa ambient pressure
Power supply
4x 1.5 V AA batteries
Battery life
For approx. 300 measurements, depending on levels of blood pressure,
pump pressure and ECG measurement length
Classification
Internal supply, IP21, no AP or APG, continuous operation
Blood pressure: Application part, type BF
ECG stick: Application part type CF
Data transfer via
Bluetooth
®
wireless technology
The product uses
Bluetooth
®
low energy technology, frequency band
2.400 – 2.483 GHz, maximum transmission power radiated in the frequen-
cy band < 20 dBm, compatible with
Bluetooth
®
≥ 4.0 smart phones /
tablets
The serial number is located on the device or in the battery compartment.
Technical information is subject to change without notification to allow for updates.
• This device conforms with the European standard EN 60601-1-2 (in accordance with CISPR 11,
IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-8) and is subject to particular precautions with regard to
electromagnetic compatibility. Please note that portable and mobile HF communication systems may
interfere with this device.
Summary of Contents for BM 96 Cardio
Page 204: ...204 1 1 1 22 42 1 USB 1 AAA 1 5 LR6 4 1 1 1 1 2 21 PAP Storage Transport...
Page 206: ...206 8...
Page 207: ...207 4...
Page 208: ...208...
Page 209: ...209 20 C 2...
Page 210: ...210...
Page 211: ...211 5 1 2 3 8 4 5 6 7 9 1 2 3 4 5 USB 6 7 M1 M2 8 9 10 11 10 11...
Page 213: ...213 6 1 5 AA LR6 Bluetooth Bluetooth 12 24 M2 M2 M2 M1...
Page 215: ...215 30 5 1 164 279 22 42...
Page 216: ...216 2 3 OK...
Page 217: ...217 60 M1 1 M2 2 30...
Page 218: ...218 4 6 BP 5 4...
Page 219: ...219...
Page 221: ...221 1 2 1 2 1 2 1 2 5...
Page 222: ...222 60 M1 1 M2 2 7 5 00 9 00 7 18 00 20 00...
Page 223: ...223 8...
Page 224: ...224 OK C A B B C...
Page 225: ...225 C A B A 2 Top Electrodes Top Electrodes 5 cm 10...
Page 226: ...226 B 3 Top Electrodes Top Electrodes 5 cm 10...
Page 227: ...227 1 C A B M1 M2 ECG 10...
Page 228: ...228 60 bpm 100 bpm...
Page 232: ...232 11 300...
Page 233: ...233 Bluetooth Bluetooth...
Page 234: ...234 12 WEEE Waste Electrical and Electronic Equipment Pb Cd Hg 13 BM 96...