LaserSpeed 8000/9000 I/O Module Instruction Handbook
Setup
Part No. 93342 / Drawing No. 0921-01516
Page 38 of 84
Revision E (Dec 2013)
Set Value
~B<B>,<S>,<D>,<P>,<M>,<E><CR>
.B<CR>
Read Settings (without labels)
Example Response Format:
230400,1,8,NONE,0,10<CR>
Where
<B>
,
<S>
,
<D>
,
<P>,
<M>
and
<E>
are as follows:
Parameter
Description
Valid Values
<B>
Baud Rate
230400, 115200, 57600,
38400, 19200, 4800
<S>
Stop Bits
1, 2
<D>
Data Bits
7, 8
<P>
Parity
Should always be set to
NONE
<M>
Sync Mode
0 or 1
<E>
Elongation Averaging
Number of elongation
readings to average
from 1 to 1000
Default 10
This command configures the baud rate and framing for the RS-422
communication link between the I/O Module and the LaserSpeed gauge. These
settings must be set to ma
tch the gauge’s communication settings for
communication between the two devices to function. The valid baud rates are
listed in the table above.
Recommended settings:
Setting
Value
Baud Rate
230400
Stop Bits
1
Data Bits
8
Parity
None
SyncMode
0
ElongAvg
10