17
2 µF
1 nF
± (1.9 % of the measured value + 8 digits)
20 µF
10 nF
± (1.9 % of the measured value + 8 digits)
200 µF
100 nF
± (1.9 % of the measured value + 8 digits)
2 mF
1 µF
± (1.9 % of the measured value + 8 digits)
< 10 digits in case of fluctuating indication
7.8 Frequency ranges
(BENNING MM 1-2/ 1-3)
Overload protection for frequency measurements: 600 V
eff
Minimum pulse width > 25 ns; operating cycle limitation > 30 % and < 70 %
Measuring
range
Resolution
Measuring accuracy for 5 V
eff
max.
Sensitivity
2 kHz
1 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
20 kHz
10 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
200 kHz
100 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
2 MHz
1 kHz
± (0.01 % of the measured value + 1 digits)
> 2 < 5 V
eff
20 MHz
10 kHz
± (0.01 % of the measured value + 1 digits)
> 2 < 5 V
eff
7.9 Temperature ranges °C
(BENNING MM 1-3)
Temperature measurements (BENNING MM 1-3) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
eff
Measuring range
Resolution
Measuring accuracy
-20 °C ~ 0 °C
1 °C
± (2 % + 4 °C)
1 °C ~ 100 °C
1 °C
± (1 % + 3 °C)
101 °C ~ 500 °C
1 °C
± (2 % + 3 °C)
501 °C ~ 800 °C
1 °C
± (3 % + 2 °C)
7.10 Temperature ranges °F
(BENNING MM 1-3)
Temperature measurements (BENNING MM 1-3) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
eff
Measuring range
Resolution
Measuring accuracy
-4 °F ~ 32 °F
1 °F
± (2 % + 8 °F)
33 °F ~ 212 °F
1 °F
± (1 % + 6 °F)
213 °F ~ 932 °F
1 °F
± (2 % + 6 °F)
933 °F ~ 1472 °F
1 °F
± (3 % + 4 °F)
8. Measuring with the BENNING MM 1‑1/ 1‑2/ 1‑3
8.1 Preparing the measurement
Operate and store the BENNING MM 1-1/ 1-2/ 1-3 at the specified storage and
operating temperatures only!
Do not permanently expose the device to sunlight.
- Check stated nominal voltage and nominal current on the safety measuring
leads. Nominal voltage and current of the enclosed safety measuring leads
comply with the respective values of the BENNING MM 1-1/ 1-2/ 1-3.
- Check insulation of the safety measuring leads. If the insulation is damaged,
the safety measuring leads must be replaced immediately.
- Check the safety measuring leads for continuity. If the conductor in the
safety measuring lead is interrupted, replace the safety measuring leads
immediately.
- Before selecting another function by means of the rotary switch
8
, discon-
nect the safety measuring leads from the measuring point.
- Strong sources of interference in the vicinity of the BENNING MM 1-1/ 1-2/ 1-3
might involve unstable readings and measuring errors.
8.2 Voltage and current measurement
Do not exceed the maximum permitted voltage with respect to
earth potential! Electrical danger!
The highest voltage which may be applied to the
- COM jack
J
02/ 2011
BENNING MM 1-1/ 1-2/ 1-3