18
- jack for
V, Ω,
, Hz
9
- jack for the 10 A range
K
(BENNING MM 1-2/ 1-3)
of the BENNING MM 1-1/ 1-2/ 1-3 against ground is 1000 V.
Electrical danger!
Maximum circuit voltage for a measured current of 500 V! In
case of fuse triggering at values above 500 V, damages of the
device might be involved. A damaged device might represent
an electrical hazard!
8.2.1 Voltage measurement
- Select the desired function (V AC) or (V DC) by means of the rotary switch
8
of the BENNING MM 1-1/ 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack
J
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for
V, Ω,
, Hz
9
of the
BENNING MM 1-1/ 1-2/ 1-3.
-
Bring the safety measuring leads into contact with the measuring
points and read the measured value on the digital display
of the
BENNING MM 1-1/ 1-2/ 1-3.
See figure 2:
DC voltage measurement
See figure 3:
AC voltage measurement
8.2.2 Current measurement (BENNING MM 1-2/ 1-3)
- Select the desired range and function (A AC) or (A DC) by means of the
rotary switch
8
of the BENNING MM 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack
J
of the
BENNING MM 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for the 10 A range
K
(DC
or AC currents up to 10 A) of the BENNING MM 1-2/ 1-3.
-
Bring the safety measuring leads into contact with the measuring
points and read the measured value on the digital display
of the
BENNING MM 1-2/ 1-3.
See figure 4:
DC current measurement (BENNING MM 1-2/ 1-3)
See figure 5:
AC current measurement (BENNING MM 1-2/ 1-3)
8.3 Resistance measurement
- Select the desired function (Ω) by means of the rotary switch
8
of the
BENNING MM 1-1/ 1-2/ 1-3
- Connect the black safety measuring lead to the COM jack
J
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V,
,
, Hz
9
of the
BENNING MM 1-1/ 1-2/ 1-3.
-
Bring the safety measuring leads into contact with the measuring
points and read the measured value on the digital display
of the
BENNING MM 1-1/ 1-2/ 1-3.
See figure 6:
Resistance measurement
8.4 Diode test
- Select the desired function (
) by means of the rotary switch
8
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack
J
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V,
,
, Hz
9
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Bring the safety measuring leads into contact with the diode connec-
tions and read the measured value on the digital display
of the
BENNING MM 1-1/ 1-2/ 1-3.
- For a standard Si diode applied in conduction direction, a conduction
voltage between 0.400 V and 0.900 V is displayed. „000“ indicates a short-
circuit inside the diode, „OL“ indicates an interruption inside the diode.
- For a diode applied in reverse direction, „OL“ is indicated. If the diode is
defective, „000“ or other values are indicated.
See figure 7:
Diode test
8.5 Continuity test with buzzer
- Select the desired function (
) by means of the rotary switch
8
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the black safety measuring lead to the COM jack
J
of the
BENNING MM 1-1/ 1-2/ 1-3.
- Connect the red safety measuring lead to the jack for V,
,
, Hz
9
of the
BENNING MM 1-1/ 1-2/ 1-3.
02/ 2011
BENNING MM 1-1/ 1-2/ 1-3