![Bender IR5002 Operating Manual Download Page 34](http://html1.mh-extra.com/html/bender/ir5002/ir5002_operating-manual_2731566034.webp)
34
TGH 1360E
Technical data
Type tests
Test of the Electromagnetic Compatibility (EMC):
Immunity for industrial environments acc. to IEC 61000-6-2:
Electrostatic discharge immunity test acc. to IEC 61000-4-2
severity degree 3
Radiated, radio-frequency, electromagnetic field
immunity test EM acc. to IEC 61000-4-3
severity degree 3
Burst immunity test acc. to IEC 61000-4-4
severity degree 3
Surge immunity test acc. to IEC 61000-4-5
severity degree 3
Immunity to conducted disturbances,
induced by radio-frequency fields acc. to IEC 61000-4-6
severity degree 3
Power frequency magnetic field
Immunity test acc. to IEC 61000-4-8
severity degree 3
Voltage dips, short interruptions and voltage
variations, immunity tests acc. to IEC 61000-4-11
severity degree 3
Emissions acc. to EN 50081-2:
Emissions acc. to EN 55011/CISPR11
class B
Mechanical tests:
Shock resistance acc. to IEC 6068-2-27
15 g/11 ms
Bumping acc. to IEC 6068-2-29
40 g/6 ms
Vibration strength acc. to IEC 6068-2-6
10...150 Hz/0.15 mm - 2g
Environmental conditions
Ambient temperature, during operation
-15
°
C...+55
°
C
Storage temperature range
-40
°
C...+70
°
C
Climatic class acc. to IEC 60721
3K5
(except condensation and formation of ice)
General data
Operating mode
continuous operation
Mounting
any position
Connection
terminals with self-lifting clamp-washers
Cross sectional area of connecting cable, single wire
0.2...4 mm
2
Cross sectional area of connecting cable, flexible
0.2...2.5 mm
2
(AWG 24 12)
Casing
19" casing 30 TE /4 HE for wall-mounting
Protection class acc to EN 60529
IP 65
Ordering details
IR5002
Art. No. 91 059 010
AG5002
Art. No.
98 039 009