23
TGH 1360E
Setting and operation
Setting the serial
interface
Select the sub menu 6 SERIAL INTERFACE from the main menu for the
configuration of the serial interfaces.
1. MAIN MENU
2. RS232 BAUD: 9600
3. RS485 BAUD: 9600
ENTER
Selecting 2 RS232 BAUD changes the baud rate of the RS232 interface. The
baud rate can be set between "No" (interface deactivated) and 19200 baud.
1. MAIN MENU
2. RS232 BAUD: 9600
3. RS485 BAUD: 9600
ENTER
By scrolling the UP and DOWN keys the desired baud rate can be selected.
Confirming the selected value with ENTER returns you to the previous menu.
In the same way the baud rate of the RS485 interface can be set.
Adjusting digital
input 1
If you are going to adjust the digital input 1 (connection terminals E1 and E6),
select the sub menu 7 ENABLE MEASUREM. from the main menu.
1. MAIN MENU
2. ENABLE MEASUREM.
INPUT: HIGH ACT.
E N T E R
After selecting the sub menu 2 ENABLE MEASUREM INPUT and confirming
with ENTER select between NOT ACTIVE, HIGH ACTIVE and LOW ACTIVE
by scrolling the UP and DOWN keys.
Explanation:
NOT ACTIVE
The external digital input 1 (E1/E6) is not activated. The
device is working independently of the signal at the
external input.
HIGH ACTIVE
A D.C. voltage of 24 V (
±
25%) activates the device. If no
DC voltage is detected the internal measuring suppression
will be activated, the measurement is disabled. The LC
display indicates MEASURING SUPPRESSION ACTIVE.
1. MAIN MENU
2. ENABLE MEASUREM.
INPUT: HIGH ACT.
E N T E R