15. Emergency inspection procedure

         Some quick inspection procedures are described below.
           1. Operation input inspection. 
                The inspection should be performed with the cable to the connector S1 disconnected, the relay p.c.board connected, and the 
                power turned on.
                Jump the S1s No. 10 pin  and the pins No. 1 to No. 8 one by one. If the p.c.board responds, the p.c.board is operating normally.
                In this case, inspect the SW.
           2. Headrest connector inspection. 
                The inspection should be performed with the cable to the connector S2 disconnected, the relay p.c.board connected, and the
                power turned on.
                Jump the S2s No. 6 pin  and the pins No. 1 to No. 4 one by one. If the p.c.board responds, the p.c.board is operating normally.
                In this case, inspect the SW.
           3. Inspection of the relay p.c.board. 
               The inspection should be performed using the harness connector, with the cable to the S10 connector disconnected.
               Jump the S10s cable connector No. 10 and pins No. 1 to No. 9 using wire and solder paste and check the relay sound, motor, and
               solenoid valve operation. The jump to the No.1 pin should be checked momentarily because the pin corresponds to the motor.
           4. Always check the potentiometer voltage. Check the voltage over the entire stroke range for smooth variation.

-7-

Summary of Contents for MCX-II

Page 1: ...MCX II FOR THE LEGREST FOLDING CHAIRS TECHNICAL INFORMATION R DENTAL CHAIR FUNCTION AND OPERATION Micro Computer Chip Version1 5 P C Board 3 E...

Page 2: ...bed separately 3 Operation sound on off options Whether or not to enable the SW operation sound is selected The change and setup procedures are described separately 4 Doctor ID options Up to four doct...

Page 3: ...ing input Backrest reclining is also prohibited Setup procedure 1 Slide the limit SW on the p c board to the limit position The red LED on the p c board turns on 2 Press and hold down the mode SW 3 Af...

Page 4: ...ds or more without any operations Pressing the mode SW again will shift the mode to the leg rest operation mode The mode is terminated automatically after a lapse of 10 seconds 5 Auto return preset an...

Page 5: ...sounding 3 Depending on the setup target press the SWs shown below Margin SW to be pressed Margin SW to be pressed Chair up Chair up Chair down Chair down Backrest raise Backrest raise Backrest recli...

Page 6: ...y to the p c board LED 3 Red Limit status indicator The LED is on while the slide SW is at the limit position LED 4 Amber Serial 3 transmission indicator The LED is turned on when the flow of transmis...

Page 7: ...up as required Operation by foot SW unavailable Check for open circuit between the SW and the p c board The identical positions set for preset auto return and rinsing slightly move in actual operatio...

Page 8: ...nected the relay p c board connected and the power turned on Jump the S2s No 6 pin and the pins No 1 to No 4 one by one If the p c board responds the p c board is operating normally In this case inspe...

Page 9: ...L Poo poo poo Poo poo poo Doctor ID setup mode Poo poo poo poo Poo poo poo poo Beeper enable disable setup mode The display shows the setup status Pip pip pip pip pip pip pip pip pip pip Leg rest oper...

Page 10: ...1 1 2 Chome Higashi shinsaibashi Chuo ku Osaka Japan TEL 06 6213 5945 FAX 06 6212 3680 TAKARA BELMONT CORPORATION NOTE Printed in Japan 3 Edition 0602 R...

Reviews: