          Manual operation: 
           The chair is operated while the SW is pressed and held down.
           1.Up       : The chair goes up while the Up SW is pressed and held down (up to its upper limit).
           2.Down  : The chair goes down while the Down SW is pressed and held down (down to its lower limit).
           3.Raise   : The backrest rises while the Raise SW is pressed and held down (up to its raise limit).
           4.Recline: The backrest reclines while the Recline SW is pressed and held down (down to its recline limit).

     This instruction manual describes the following switches (referred to as SW hereafter).
      1. Limit SW : The red slide switch on the p.c.board, closer to the p.c.board center as the limit position.
      2. Store SW : The push button SW to the left under the indicator.
      3. Mode SW : The push button SW to the right under the indicator.
      4. JP1           : The jumper terminal on the p.c.board.

   

1. The options available for each function are described below.

       The following options are available, depending on the model used.
           1. Last-position operation options: 
               The operation varies depending on the JP1 configuration on the p.c.board.
                  1-2 short: Backrest movement only.
                  2-3 short: Backrest and up/down operations.
           2. P3/Last position function options: 
               Whether to use the LP SW as the P3 switch or last position switch is selected.
               The change and setup procedures for this programmable switch are described separately.
           3. Operation sound on/off options: 
               Whether or not to enable the SW operation sound is selected. The change and setup procedures are described separately.
           4. Doctor ID options: 
               Up to four doctor IDs can be registered for an associate dental clinic. Preset (P1 to P3) for up to four doctors are available.
               Auto return and rinsing positions are shared. The modification and setup procedures are described separately.

   2. Basic operations

           Prohibit all input: 
           The input stops all the operations of the chair.
           If an operation SW is turned on while the input is present, the alarm will sound while the SW is on.
           The Prohibit all input is assigned to the No. 3 and No. 4 pins of the S4 connector and enabled by setting the jumpers.

           Prohibit lowering input: 
           This input stops lowering, full reclining, and auto operations. 
           If the operation SWs above are turned on while the input is present, the alarm will sound while the SW is on.
           This input is assigned to the No. 1 and No. 2 pins of the S4 connector, and enabled by setting the jumpers.

LS

LIMIT

NORMAL

MCX-04/01TI

sw1 sw2

LP

.SEA

T

Store Switch

Mode Switch

Limit Switch

Jumper Terminal

1

3

Belmont

FRONT

BACK

          Auto operation: 
           Once the SW is pressed, the operation will continue after the SW is released.
           When Auto and manual operations are activated simultaneously, the manual operation will over-ride.
           1. Auto Return (P0)
               Chair returns to the initial preset position.
           2. P1, P2
               Chair moves to a preset position.
               The leg rest will also move to the preset position. (Appended June 28, 2005)

-1-

Summary of Contents for MCX-II

Page 1: ...MCX II FOR THE LEGREST FOLDING CHAIRS TECHNICAL INFORMATION R DENTAL CHAIR FUNCTION AND OPERATION Micro Computer Chip Version1 5 P C Board 3 E...

Page 2: ...bed separately 3 Operation sound on off options Whether or not to enable the SW operation sound is selected The change and setup procedures are described separately 4 Doctor ID options Up to four doct...

Page 3: ...ing input Backrest reclining is also prohibited Setup procedure 1 Slide the limit SW on the p c board to the limit position The red LED on the p c board turns on 2 Press and hold down the mode SW 3 Af...

Page 4: ...ds or more without any operations Pressing the mode SW again will shift the mode to the leg rest operation mode The mode is terminated automatically after a lapse of 10 seconds 5 Auto return preset an...

Page 5: ...sounding 3 Depending on the setup target press the SWs shown below Margin SW to be pressed Margin SW to be pressed Chair up Chair up Chair down Chair down Backrest raise Backrest raise Backrest recli...

Page 6: ...y to the p c board LED 3 Red Limit status indicator The LED is on while the slide SW is at the limit position LED 4 Amber Serial 3 transmission indicator The LED is turned on when the flow of transmis...

Page 7: ...up as required Operation by foot SW unavailable Check for open circuit between the SW and the p c board The identical positions set for preset auto return and rinsing slightly move in actual operatio...

Page 8: ...nected the relay p c board connected and the power turned on Jump the S2s No 6 pin and the pins No 1 to No 4 one by one If the p c board responds the p c board is operating normally In this case inspe...

Page 9: ...L Poo poo poo Poo poo poo Doctor ID setup mode Poo poo poo poo Poo poo poo poo Beeper enable disable setup mode The display shows the setup status Pip pip pip pip pip pip pip pip pip pip Leg rest oper...

Page 10: ...1 1 2 Chome Higashi shinsaibashi Chuo ku Osaka Japan TEL 06 6213 5945 FAX 06 6212 3680 TAKARA BELMONT CORPORATION NOTE Printed in Japan 3 Edition 0602 R...