background image

     14. If a failure is suspected, the following inspection should be conducted.

Symptom

Inspection target

Troubleshooting

The beeper sounds continuously and 
the chair will not move at limit setup.

Measure the potentiometer voltage.The 
operation is prohibited around 0 V or 5 V, 
which are potentiometer limits.

Reinstall the potentiometer to the chair.

The beeper sounds continuously at limit 
setup whereas the chair is operable.

The wiring to the potentiometer may be 
improper. Check the no. 1 and 3 lines. If 
the symptom is momentary, the condition 
is normal.

If the lines are connected opposite to each 
other, correct the connection according to 
the wiring specifications.

The chair will not move.

LED 2 is off.

Check for power supply.

The chair will not move.

LED 1 is off.

Check if prohibit all input is present or not.

Operation from the unit unavailable.

Check the transmission LED on the unit.

If the LED on the unit is off, the malfunction 
is caused by poor connection or open 
circuit.If the LED on the unit is on, check 
for baud-rate consistency and set up as 
required.

Operation by foot SW unavailable.

Check for open circuit between the SW 
and the p.c.board.

The identical positions set for preset, 
auto-return, and rinsing slightly move in 
actual operation.

Check the margins.

Increase the margins.

The preset position is not reached.

Check the margin.

Decrease the margin.

The preset position is exceeded.

Check the margin.

Increase the margin.

The chair is not moved by pressing a 
SW.The beeper sounds while the SW is 
pressed.

An input to prohibit the chair operation is 
present. Cancel the  input to confirm the 
kind of prohibit signal.

Cancel the prohibit input.

The beeper sounds at power on and 
stops when a SW is pressed. The auto 
operation is disabled thereafter, and 
the manual operation works only in one 
direction.

Open or short circuit of the potentiometer. 
Measure the voltage using a multimeter 
to check for the malfunction.

Repair the open circuit, replace the cable.

The stop position of the chair becomes  The position of the potentiometer is 

Check for loose installation of the device 

The legrest does not extend by preset 
switches.

Check the set up position of legrest.

Manually extend the lrgrest for  setting-up. 
Extend the legrest in independent mode to 
the limit position and set the preset.

-6-

Summary of Contents for MCX-II

Page 1: ...MCX II FOR THE LEGREST FOLDING CHAIRS TECHNICAL INFORMATION R DENTAL CHAIR FUNCTION AND OPERATION Micro Computer Chip Version1 5 P C Board 3 E...

Page 2: ...bed separately 3 Operation sound on off options Whether or not to enable the SW operation sound is selected The change and setup procedures are described separately 4 Doctor ID options Up to four doct...

Page 3: ...ing input Backrest reclining is also prohibited Setup procedure 1 Slide the limit SW on the p c board to the limit position The red LED on the p c board turns on 2 Press and hold down the mode SW 3 Af...

Page 4: ...ds or more without any operations Pressing the mode SW again will shift the mode to the leg rest operation mode The mode is terminated automatically after a lapse of 10 seconds 5 Auto return preset an...

Page 5: ...sounding 3 Depending on the setup target press the SWs shown below Margin SW to be pressed Margin SW to be pressed Chair up Chair up Chair down Chair down Backrest raise Backrest raise Backrest recli...

Page 6: ...y to the p c board LED 3 Red Limit status indicator The LED is on while the slide SW is at the limit position LED 4 Amber Serial 3 transmission indicator The LED is turned on when the flow of transmis...

Page 7: ...up as required Operation by foot SW unavailable Check for open circuit between the SW and the p c board The identical positions set for preset auto return and rinsing slightly move in actual operatio...

Page 8: ...nected the relay p c board connected and the power turned on Jump the S2s No 6 pin and the pins No 1 to No 4 one by one If the p c board responds the p c board is operating normally In this case inspe...

Page 9: ...L Poo poo poo Poo poo poo Doctor ID setup mode Poo poo poo poo Poo poo poo poo Beeper enable disable setup mode The display shows the setup status Pip pip pip pip pip pip pip pip pip pip Leg rest oper...

Page 10: ...1 1 2 Chome Higashi shinsaibashi Chuo ku Osaka Japan TEL 06 6213 5945 FAX 06 6212 3680 TAKARA BELMONT CORPORATION NOTE Printed in Japan 3 Edition 0602 R...

Reviews: