Installation
14
METPOINT® BDL
10 Installation
10.1 Safety instructions
Danger!
Supply voltage!
The contact with non-insulated parts carrying supply voltage involves the risk of an electric shock
resulting in injuries and death.
Measures:
•
Observe all regulations in effect during the electrical installation (e.g. VDE 0100)!
•
Any electrical works must only be carried out by authorised and skilled personnel .
•
The connection to the electric mains and the corresponding protective equipment must comply with the
legal provisions in force at the place of installation of the
METPOINT
®
BDL
; the connection and installa-
tion must be carried out by specialists who are qualified accordingly.
•
Make sure that no parts of the measuring devices are energized and that the measuring devices cannot
be connected with the electric supply mains during service measures.
Danger!
Missing earth connection!
When the earth connection (protective earth) is missing, there is the risk in the event of
malfunction, that contactable, conductive components may carry supply voltage. Touching
of such parts leads to an electric shock with injuries and death.
It is imperative to connect the plant to earth or to connect the protective conductor accord-
ing to the regulations.
Do not use plug adapters at the power plug.
Have the power plug replaced by qualified skilled persons, if required.
Danger!
Missing separator!
All voltages representing a hazard when coming into contact with them must be discon-
nectable via assigned separators that need to be installed externally.
The separator must be installed in the vicinity of the device.
The separator needs to correspond to IEC 60947-1 and IEC 60947-3.
The separator must separate all voltage-carrying conductors.
The separator must not be installed in the power supply line.
The separator must be easily accessible for the user.