Technical data BDL
METPOINT® BDL
11
Pos : 24 /Beko T ec hnis che D okumentati on/Übersc hriften/1/T ec hni sche D aten @ 0\mod_1184329570967_16746.doc x @ 16779 @ 1 @ 1
8.1
Input signals
ten/1/Maßzeic hnung @ 0\mod_1183638072605_16746.doc x @ 16757 @ 1 @ 1
8.2
Line cross-sections
5.1 Voltage supply 100 – 240 VAC, 50 – 60 Hz, special version 24 VDC:
Cable cross-section power supply:
0.75 mm ²
5.2 Sensor connections/output signals:
Cable cross-section sensor supply:
Clamping range 3-7 mm
Input signals
Signal current
(0 – 20 mA/4 – 20 mA)
Internal or external voltage supply
Measuring range
0 – 20 mA/4 – 20 mA
Resolution
0.0001 mA
Accuracy
±
0.003 mA
±
0.05%
Input resistance
50
Ω
Signal voltage
(0 – 1 V)
Measuring range
0 – 1 V
Resolution
0.05 mV
Accuracy
±
0.2 mV
±
0.05%
Input resistance
100 k
Ω
Signal voltage
(0 – 10 V/30 V)
Measuring range
0 – 10 V/30 V
Resolution
0.5 mV
Accuracy
±
2 mV
±
0.05%
Input resistance
1 M
Ω
RTD
Pt100
Measuring range
-200 – 850°C
Resolution
0.1 °C
Accuracy
±
0.2°C at -100 – 400°C
±
0.3 °C (remaining range)
RTD
Pt1000
Measuring range
-200 – 850°C
Resolution
0.1°C
Accuracy
±
0.2°C at -100 – 400°C
±
0.3°C (remaining range)
Pulse
Measuring range
Min. pulse duration 100 µS
Frequency 0 – 1 kHz
Max. 30 VDC