background image

16

When the required number of break-points has
been entered, return to the linearisation sub-menu
by pressing 

E

.  

The indicator will display the 'Add'

or 'dEL' prompt depending upon the last sub-
function used.  The indicator input current and
corresponding indicator display at each break-
point, can now be entered using the ‘in’ and ‘diSP’
sub-functions.

Using the 

 or 

 button select 'in' from the sub-

menu and press 

P

 which will reveal the first break-

point  '0 : n', where n is the total number of break-
points entered. Press 

P

 and use the 

 and 

buttons and the 

P

 button to move between digits,

to enter the input current at which the first break-
point is required, usually 4.000mA.   When set as
required, press 

E

 to return to the ‘0  :  n’ prompt

from which the the next break-point can be
selected using the 

 and 

 buttons.  When the

required break-point has been selected press 

P

and enter the indicator input current at which this
break-point is required. Repeat this procedure until
the indicator input current at all the break-points
has been defined and then return to the ‘in’ sub-
function by pressing the 

E

 button.

The corresponding indicator display at each of the
break-points can now be defined with the ‘diSP’
sub-function    Using the 

 and 

 buttons select

the ‘diSP’ sub-function and press 

P

 which will

reveal the first break-point '0 : n', where n is the
total number of break-points entered. Press 

P

 and

use the 

 and 

 buttons and the 

P

 button to

move between digits, to enter the required
indicator display at this first break-point. When set
as required, press 

E

 to return to the ‘0  :  n’ prompt

from which the the next break-point can be
selected using the 

 or 

 buttons.  When the

required break-point has been selected press 

P

and set the required indicator display at this
break-point.

Repeat this procedure until the indicator display at
all the break-points has been defined and then
return to the ‘SEt’ function in the configuration
menu by pressing the 

E

 button twice.

6.3  Lineariser error message

If an attempt is made to position a break-point at a
current which is not greater than the current of the
preceeding break-point, or at a current which is not
less than the current of the following break-point,
the error message ‘FaiL’ will be displayed.  This
error message will also be displayed if an attempt
is made to position a  break-point outside the
current range 3.8 to 21.0mA.

6.4 Under and over-range

The lineariser does not change the under and
over-range indication described in section 5.12.
At input currents below that specified for the first
break-point (0), the indicator will continue to use
the specified slope of the first segment.  Although
not guaranteed, most indicators will continue to
function normally with an input current between
1.8mA and 4mA, at lower currents the instrument
will display ‘LPLo’ before it stops functioning.

At input currents above that specified for the last
break-point, the indicator will continue to use the
slope specified for the last lineariser segment.

6.5  Lineariser default configuration

When the lineariser is reset to the default
conditions using the ‘rSEt’ reset function described
in section 5.11, the defaults conditions are:

BA507E  BA527E
BA508E  BA528E

First break-point 4mA

0.0

  0.00

Second break-point 20mA

100.0

100.00

Summary of Contents for BA507E

Page 1: ...Supplied by Call us on 44 0 118 916 9420 Email info 247able com com Installation Maintenance Instructions BA507E BA508E BA527E BA528E...

Page 2: ...rk Danehill Lower Earley Reading Berkshire RG6 4UT UK Tel 44 0 118 9311188 Email info able co uk Aberdeen Office Unit 6 Airside Business Park Kirkhill Industrial Estate Dyce Aberdeen AB21 0GT UK Tel 4...

Page 3: ...Issue 3 1st December 2012 BA507E BA527E BA508E BA528E General Purpose loop powered panel mounting indicators issue 3...

Page 4: ...error message 6 4 Lineariser under and over range 6 5 Lineariser default configuration 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Rout...

Page 5: ...splay any variable represented by the 4 20mA current e g temperature flow pressure or level The BA507E BA527E BA508E and the BA528E are not certified for use in explosive atmospheres For hazardous are...

Page 6: ...in display mode function has been enabled P E Access to configuration menu via optional security code Note BA527E and BA528E only If the indicator has been calibrated using the CAL function calibratio...

Page 7: ...el providing that the operating temperature is between 40 C and 70 C Figs 4A and 4B show the overall dimensions of the 96 x 48mm and the 144 x 72mm instruments together with the recommended panel cut...

Page 8: ...dovetail as shown in Fig 5 Push the knurled screw slightly forward to engage the thread and tighten by turning clockwise until it is just finger tight When both clamps are fitted ensure that the gaske...

Page 9: ...rd unclip the protruding end of the flexible strip by gently pushing it upwards and pulling it out of the enclosure Peel the existing scale card from the flexible strip and replace it with a new print...

Page 10: ...00 0000 Function FunC Std Std Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00 Bargraph finish BarHi 100 00 P button in display...

Page 11: ...value Using ZEro function set required display at 4mA by pressing pressing and to move to the next digit Similarly using SPAn function set required display at 20mA E P P E P E P E E Select type of ba...

Page 12: ...ble lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or press the or button to...

Page 13: ...ese conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrument will be reverse acting i e an increasi...

Page 14: ...between on and oFF When set as required press E to return to the tARE prompt in the configuration menu 5 10 Security code CodE Access to the instrument configuration menu may be protected by a four di...

Page 15: ...lling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break points required should first be entered using t...

Page 16: ...14...

Page 17: ...a new break point before the displayed break point The calibration of existing break points is not changed but the identification number of all subsequent break points is increased by one dEL Remove a...

Page 18: ...ired press E to return to the 0 n prompt from which the the next break point can be selected using the or buttons When the required break point has been selected press P and set the required indicator...

Page 19: ...s does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop will be...

Page 20: ...output may be independently conditioned as a high or low alarm with a normally open or normally closed output in the non alarm condition When the 4 20mA current powering the indicator is removed both...

Page 21: ...19...

Page 22: ...section 8 3 3 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator display equals the setpoint See section 8 3 4 Hi Lo Alarm function Defines the alarm function a...

Page 23: ...or fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Page 24: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Page 25: ...plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 15 Fig 15 Displayed value and setpoints on bargraph T...

Page 26: ...backlight 8 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the indicator 4 20mA input as shown in Fig 17 which increases the maximum indicator voltage dr...

Reviews: