background image

12

The indicator’s digital display at which the
bargraph starts is defined by the ‘bArLo’ sub-
function which is selected by pressing the 

button followed by the 

P

 button which will reveal

the current indicator display at which the bargraph
starts.   The flashing digit can be adjusted by
pressing the 

 or 

 buttons, when set as required

pressing 

P

 

will transfer control to the next digit.

When all the digits have been adjusted, press 

E

 to

return to the 'bArLo' prompt from which ‘bArHi’
which defines the finishing point of the bargraph
can be selected by pressing the 

 button.   ‘bArHi’

is adjusted in the same way as ‘bArLo’.   When set
as required, pressing 

E

 twice will return the display

to the ‘bAr’ prompt in the configuration menu.

Note:

 ‘bArLo’ must be set lower than ‘bArHi’,

incorrect setting is indicated by the bargraph scale
flashing with a single bargraph segment activated.

5.8  Function of the 

P

 push-button:   ‘C - - P’

When the indicator is in the display mode,
operating the 

P

 push-button will display the input

current in milliamps, or the displayed value as a
percentage of the difference between the
displayed values at 4mA and 20mA inputs.

To check or change the function of the 

P

 push-

button select 'C - -P' from the configuration menu
and press 

P

 to reveal the current setting.  Pressing

the 

 or 

 button will toggle the setting between

'4-20' the current display in milliamps and 'PC' the
percentage display.   When set as required press

E

 to return to the ‘C - - P’ prompt in the

configuration menu.

5.9  Tare function:   ‘tArE’

The tare function is primarily intended for use with
weighing system.  When the indicator is in the
display mode and the tare function is activated,
pressing the 

E

 button for more than three seconds

will zero the indicator’s digital display and activate
the tare annunciator.  On models with a bargraph,
the bargraph remains linked to the digital display
when the tare function is activated.   Subsequent
operation of the 

E

 push-button for less than 3

seconds will return the indicator to the gross
display and deactivate the tare annunciator.

To check or change the tare function select 'tARE'
from the configuration menu and press 

P

 to reveal

the current setting.  Pressing the 

 or 

 button

will toggle the setting between 'on' and 'oFF'.
When set as required press 

E

 to return to the

‘tARE’ prompt in the configuration menu.

5.10  Security code:  ‘CodE’

Access to the instrument configuration menu may
be protected by a four digit security code which
must be entered to gain access.   New instruments

are configured with the default security code 0000
which allows unrestricted access to all
configuration functions.

To enter a new security code select 'CodE' from
the configuration menu and press

 

P

 which will

cause the indicator to display the existing security
code with one digit flashing.  The flashing digit can
be adjusted using the 

 and

 

 push-buttons,

when set as required operating the 

P

 button will

transfer control to the next digit.   When all the
digits have been adjusted press

 

E

 to return to the

‘CodE’ prompt. The revised security code will be
activated when the indicator is returned to the
display mode.  Please contact BEKA associates
sales department if the security code is lost.

5.11  Reset to factory defaults:  ‘rSEt’

This function enables the indicator and the
lineariser to be quickly returned to the factory
default configurations shown in sections 5 and 6.

To reset the indicator or lineariser select ‘rSEt’
from the configuration menu and press 

P

, the

indicator will display one of the reset options
‘ConF’ or ‘LtAb’.

‘ConF’

Resets the indicator to defaults

‘LtAb’

Resets the lineariser to defaults

Using the 

 or 

 push-button select the required

sub-function and press 

P.

  To prevent accidental

resetting the request must be confirmed by
entering ‘5urE’.   Using the 

 button set the first

flashing digit to ‘5’ and press 

P

 to transfer control

to the second digit which should be set to ‘u’.
When ‘5urE’ has been entered pressing the 

E

button will reset the selected configuration menus
and return the display to the ‘rSEt’ function in the
configuration menu.

5.12  Under and over-range

If the numerical display range of the indicator is
exceeded, all the decimal points will flash as
shown below:

BA507E

 BA527E

BA508E

 BA528E

Underrange

-9.9.9.9

-9.9.9.9.9

Overrange

 9.9.9.9

 9.9.9.9.9

Although not guaranteed, most indicators will
continue to function normally with an input current
between 1.8mA and 4mA, at lower currents the
instrument will display ‘LPLo’ before it stops
functioning.

Under or overrange of the BA527E and BA528E
bargraph is indicated by an activated arrow at the
appropriate end of the bargraph and a flashing
bargraph scale.

Summary of Contents for BA507E

Page 1: ...Supplied by Call us on 44 0 118 916 9420 Email info 247able com com Installation Maintenance Instructions BA507E BA508E BA527E BA528E...

Page 2: ...rk Danehill Lower Earley Reading Berkshire RG6 4UT UK Tel 44 0 118 9311188 Email info able co uk Aberdeen Office Unit 6 Airside Business Park Kirkhill Industrial Estate Dyce Aberdeen AB21 0GT UK Tel 4...

Page 3: ...Issue 3 1st December 2012 BA507E BA527E BA508E BA528E General Purpose loop powered panel mounting indicators issue 3...

Page 4: ...error message 6 4 Lineariser under and over range 6 5 Lineariser default configuration 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Rout...

Page 5: ...splay any variable represented by the 4 20mA current e g temperature flow pressure or level The BA507E BA527E BA508E and the BA528E are not certified for use in explosive atmospheres For hazardous are...

Page 6: ...in display mode function has been enabled P E Access to configuration menu via optional security code Note BA527E and BA528E only If the indicator has been calibrated using the CAL function calibratio...

Page 7: ...el providing that the operating temperature is between 40 C and 70 C Figs 4A and 4B show the overall dimensions of the 96 x 48mm and the 144 x 72mm instruments together with the recommended panel cut...

Page 8: ...dovetail as shown in Fig 5 Push the knurled screw slightly forward to engage the thread and tighten by turning clockwise until it is just finger tight When both clamps are fitted ensure that the gaske...

Page 9: ...rd unclip the protruding end of the flexible strip by gently pushing it upwards and pulling it out of the enclosure Peel the existing scale card from the flexible strip and replace it with a new print...

Page 10: ...00 0000 Function FunC Std Std Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00 Bargraph finish BarHi 100 00 P button in display...

Page 11: ...value Using ZEro function set required display at 4mA by pressing pressing and to move to the next digit Similarly using SPAn function set required display at 20mA E P P E P E P E E Select type of ba...

Page 12: ...ble lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or press the or button to...

Page 13: ...ese conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrument will be reverse acting i e an increasi...

Page 14: ...between on and oFF When set as required press E to return to the tARE prompt in the configuration menu 5 10 Security code CodE Access to the instrument configuration menu may be protected by a four di...

Page 15: ...lling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break points required should first be entered using t...

Page 16: ...14...

Page 17: ...a new break point before the displayed break point The calibration of existing break points is not changed but the identification number of all subsequent break points is increased by one dEL Remove a...

Page 18: ...ired press E to return to the 0 n prompt from which the the next break point can be selected using the or buttons When the required break point has been selected press P and set the required indicator...

Page 19: ...s does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop will be...

Page 20: ...output may be independently conditioned as a high or low alarm with a normally open or normally closed output in the non alarm condition When the 4 20mA current powering the indicator is removed both...

Page 21: ...19...

Page 22: ...section 8 3 3 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator display equals the setpoint See section 8 3 4 Hi Lo Alarm function Defines the alarm function a...

Page 23: ...or fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Page 24: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Page 25: ...plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 15 Fig 15 Displayed value and setpoints on bargraph T...

Page 26: ...backlight 8 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the indicator 4 20mA input as shown in Fig 17 which increases the maximum indicator voltage dr...

Reviews: