background image

7

5.  CONFIGURATION AND CALIBRATION

Both indicators are configured and calibrated via
the four push buttons which are located behind the
control cover.   All the configuration functions are
contained in an easy to use intuitive menu that is
shown diagrammatically in Fig 6.

Each menu function is summarised in section 5.1
and includes a reference to more detailed
information. When the indicator is fitted with alarms
additional functions are added to the menu which
are described in section 8.3

Throughout this manual push-buttons are shown
as 

P

 E

,

  

 

or 

, and legends displayed by the

indicator are shown within inverted commas e.g.
'CAL' and ' ALr2'.

Access to the configuration menu is obtained by
operating the 

P

 and 

E

 push buttons

simultaneously.  If the indicator security code is set
to the default 0000 the first parameter 'FunC' will
be displayed.  If a security code other than the
default code 0000 has already been entered, the
indicator will display 'CodE'.  Pressing the 

P

 button

will clear this prompt allowing each digit of the
code to be entered using the 

 and

 

 push

buttons and the 

P

 button to move control to the

next digit.  When the correct four digit code has
been entered pressing

 

E

 will cause the first

parameter 'FunC' to be displayed.   If the code is
incorrect, or a button is not pressed within twenty
seconds, the indicator will automatically return to
the display mode.

Once within the configuration menu the required
parameter can be reached by scrolling through the
menu using the 

 and 

 

push buttons as shown

in Fig 6.    When returning to the display mode
following recalibration or a change to any function,
the indicator will display  ‘dAtA’ followed by ‘SAVE’
while the new information is stored in non-volatile
memory.

All new indicators are supplied calibrated as
requested at the time of ordering.  If calibration is
not requested, indicators will be supplied with the
following default configuration:

Default Configuration

BA504E BA524E

Access code  ‘CodE’

  0000

  0000

Function  ‘FunC’

Linear

Linear

Display at 4mA  ‘Zero’

0.0

0.00

Display at 20mA  ‘SPAn’

100.0

100.00

Resolution  ‘rESn’

1 digit

1 digit

Bargraph start  ‘BarLo’

-----

0.00

Bargraph finish  ‘BarHi’

-----

100.00

P

 button in display mode ‘C—P’

%

%

Tare  ‘tArE’

Off

Off

5

.1  Summary of configuration functions

This section summarises each of the main
configuration functions and includes a cross
reference to a more detailed description.  Fig 6
illustrates the location of each function within the
configuration menu.  The lineariser and the
optional factory fitted alarms are described
separately in sections 6 and 8.3 of this manual.

  

Display

      Summary of function

'FunC'

Indicator function

Defines the relationship between the
4/20mA input current and the indicator
display.  May be set to:

 

‘Std’

Standard linear relationship

‘root’

Square root extraction

‘Lin’

16 segment adjustable
lineariser – see section 6.

See section 5.2

'rESn'

Display resolution

Defines the resolution of the least
significant display digit.  May be set to
‘1’, ‘2’, ‘5’ or ‘10’ digits.
See section 5.3

'dP'

Decimal point

Positions a dummy decimal point
between any of the digits or turns it off.
See section 5.4

'CAL'

Calibration of the digital display
using an external current source.

Enables the zero and span of the
indicator to be adjusted using an
external current source such as a
calibrator.  When used with an accurate
traceable current source this is the
preferred method of calibration.
See section 5.5

'SEt'

Calibration of display using internal
references.

Enables the zero and span of the
indicator to be adjusted without the
need for an accurate input current or
disconnection from the 4/20mA loop.
See section 5.6

'bAr'

Bargraph format and calibration

Only the BA524E has a bargraph

The bargraph may be conditioned to
start from left, right or centre of the
display, or it may be disabled.  When
optional alarms are fitted it can also
display both alarm setpoints and the
measured value.
The bargraph may be calibrated to start
and finish at any value within the
indicator’s calibrated digital display.
See section 5.7

Summary of Contents for BA504E

Page 1: ...Issue 2 14th December 2011 BA504E BA524E General Purpose loop powered field mounting indicators Issue 2...

Page 2: ...ver range 6 5 Lineariser default configuration 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Custom...

Page 3: ...e g temperature flow pressure or level The BA504E and BA524E are not certified for use in explosive atmospheres For hazardous area application the BA304E BA304NE BA324E or the BA324NE should be used 2...

Page 4: ...indicator and the ACSP access setpoints in display mode function has been enabled P E Provides access to the configuration menu via optional security code Note BA524E only If the indicator has been c...

Page 5: ...ally joined and may be used for linking cable screens The indicator s earth terminal is connected to the carbon loaded GRP enclosure If this enclosure is not bolted to an earthed post or structure the...

Page 6: ...EMC Both instruments compy with the requirements of the European EMC Directive 2004 108 EC For specified immunity all wiring should be in screened twisted pairs with the screens connected at one poin...

Page 7: ...tion Default Configuration BA504E BA524E Access code CodE 0000 0000 Function FunC Linear Linear Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph...

Page 8: ...8...

Page 9: ...n 16 segment adjustable lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or pre...

Page 10: ...etween the two currents is greater than 4mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the...

Page 11: ...n and oFF When set as required press E to return to the tARE prompt in the configuration menu 5 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit secur...

Page 12: ...r volumetric units by filling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break point required should f...

Page 13: ...13...

Page 14: ...dd a break point Adds a new break point before the displayed break point The calibration of existing break points is not changed but the identification number of all subsequent break point is increase...

Page 15: ...tor display at the first break point When set as required press E to return to the 0 n prompt from which the the next break point can be selected using the or buttons When the required break point has...

Page 16: ...If this does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop w...

Page 17: ...n Both indicators can also be supplied with a blank or custom laser engraved stainless steel plate secured by two screws to the front of the instrument enclosure This plate can accommodate 1 row of 9...

Page 18: ...polarised and current will only flow in one direction Ron less than 5 0 7V Roff greater than 1M Fig 10 Equivalent circuit of each alarm output The solid state output of each alarm may be used to switc...

Page 19: ...19...

Page 20: ...setpoint The alarm is activated when the indicator display equals the setpoint See section 8 3 4 Hi Lo Alarm function Defines the alarm function as High or Low See section 8 3 5 no nC Normally open or...

Page 21: ...r fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Page 22: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Page 23: ...int plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 14 Fig 14 Displayed value and setpoints on bargra...

Page 24: ...cting it in series with the indicator s 4 20mA input as shown in Fig 16 which increases the maximum indicator voltage drop from 1 2 to 5V Fig 16 Loop powered backlight Providing the increased voltage...

Page 25: ...l pipe BA392D Stainless steel bracket secured by two worm drive hose clips Will clamp to any vertical or horizontal pipe with an outside diameter between 60 and 80mm BA393 Heavy duty stainless steel b...

Reviews: