background image

12

6

.  LINEARISER

A sixteen segment, seventeen breakpoint (0 to 16)
lineariser may be selected in the ‘FunC’ section of
the configuration menu.   The starting point and
slope of each straight line segment are fully
adjustable allowing the indicator to display most
non-linear process variables in linear engineering
units. Each break-points must occur at a current
greater than the preceeding break-point and less
than the following break-point, in the range 3.8 to
21.0mA    Fig 8 shows a typical linearised indicator
characteristic.     Fig 7 shows a typical linearised
indicator characteristic.

Fig 7 shows a typical linearising characteristic

Selecting ‘Lin’ in the ‘FunC’ section of the
configuration menu activates the lineariser, this
does not change the configuration menu shown in
Fig 6, but the 'CAL' and 'SEt' functions are
extended as shown in Fig 8.    As with a linear
indicator, calibration may be performed with an
external current source using the 'CAL' function, or
with the internal reference using the 'SEt' function.

The lineariser configuration is retained irrespective
of how the the indicator function ‘FunC’ is
subsequently changed.   It is therefore possible to
select and deselect the lineariser without having to
reconfigure it each time.

6.1  Lineariser calibration using an external

current source.

This method allows direct calibration of the
lineariser with an exteral current source and is the
preferred method when traceability is required.   If
the exact system non-linearity is unknown, this
method also allows direct calibration from the
variable to be displayed.  e.g. the output from a

level sensor in an irregular tank may be displayed
in linear volumetric units by filling the tank with
known incremental volumes and calibrating the
indicator to display the sum of the increments at
each break-point.

The number of break-point required should first be
entered using the 'Add' and 'dEL'  functions.  In
both these sub-functions the indicator initially
displays the current break-point and the total
number of break-points being used as shown
below.

Display      Description of function

'Add'

Add a break-point

Adds a new break-point before the
displayed break-point.  The calibration
of existing break-points is not changed,
but the identification number of all
subsequent break-points is increased
by one.

'dEL'

Remove a break-point

Removes the displayed break-point and
joins the preceding break-point to the
following break-point with a straight
line.   The identification number of all
subsequent break-points is decreased
by one.

To add a break-point use the 

 or 

 button to

select 'CAL' from the configuration menu and press

P

 

which will result in the

 

'Add' sub-function prompt

being displayed. To enter the sub-function press 

P

which will reveal the current break-point and the
total number of break-points which have already
been entered.    Each subsequent operation of the

P

  push-button will introduce an additional break-

point up to the maximum of 17 (breakpoint 16).
When adding a break-point to a calibrated
indicator, the insertion position for the new
segment can be selected using the 

 and 

push-buttons.

The delete break-point sub-function 'dEL' operates
in exactly the same way as the 'Add' sub-function
described above.   Once within the ‘dEL’ sub-
function each time the 

P

 button is pressed a break-

point is removed. When deleting a break-point
from a calibrated indicator, the break-point to be
deleted can be selected using the 

 and 

 push-

buttons.  The minimum number of break-points is
2, break points 0 and 1.

Summary of Contents for BA504E

Page 1: ...Issue 2 14th December 2011 BA504E BA524E General Purpose loop powered field mounting indicators Issue 2...

Page 2: ...ver range 6 5 Lineariser default configuration 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Custom...

Page 3: ...e g temperature flow pressure or level The BA504E and BA524E are not certified for use in explosive atmospheres For hazardous area application the BA304E BA304NE BA324E or the BA324NE should be used 2...

Page 4: ...indicator and the ACSP access setpoints in display mode function has been enabled P E Provides access to the configuration menu via optional security code Note BA524E only If the indicator has been c...

Page 5: ...ally joined and may be used for linking cable screens The indicator s earth terminal is connected to the carbon loaded GRP enclosure If this enclosure is not bolted to an earthed post or structure the...

Page 6: ...EMC Both instruments compy with the requirements of the European EMC Directive 2004 108 EC For specified immunity all wiring should be in screened twisted pairs with the screens connected at one poin...

Page 7: ...tion Default Configuration BA504E BA524E Access code CodE 0000 0000 Function FunC Linear Linear Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph...

Page 8: ...8...

Page 9: ...n 16 segment adjustable lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or pre...

Page 10: ...etween the two currents is greater than 4mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the...

Page 11: ...n and oFF When set as required press E to return to the tARE prompt in the configuration menu 5 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit secur...

Page 12: ...r volumetric units by filling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break point required should f...

Page 13: ...13...

Page 14: ...dd a break point Adds a new break point before the displayed break point The calibration of existing break points is not changed but the identification number of all subsequent break point is increase...

Page 15: ...tor display at the first break point When set as required press E to return to the 0 n prompt from which the the next break point can be selected using the or buttons When the required break point has...

Page 16: ...If this does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop w...

Page 17: ...n Both indicators can also be supplied with a blank or custom laser engraved stainless steel plate secured by two screws to the front of the instrument enclosure This plate can accommodate 1 row of 9...

Page 18: ...polarised and current will only flow in one direction Ron less than 5 0 7V Roff greater than 1M Fig 10 Equivalent circuit of each alarm output The solid state output of each alarm may be used to switc...

Page 19: ...19...

Page 20: ...setpoint The alarm is activated when the indicator display equals the setpoint See section 8 3 4 Hi Lo Alarm function Defines the alarm function as High or Low See section 8 3 5 no nC Normally open or...

Page 21: ...r fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Page 22: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Page 23: ...int plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 14 Fig 14 Displayed value and setpoints on bargra...

Page 24: ...cting it in series with the indicator s 4 20mA input as shown in Fig 16 which increases the maximum indicator voltage drop from 1 2 to 5V Fig 16 Loop powered backlight Providing the increased voltage...

Page 25: ...l pipe BA392D Stainless steel bracket secured by two worm drive hose clips Will clamp to any vertical or horizontal pipe with an outside diameter between 60 and 80mm BA393 Heavy duty stainless steel b...

Reviews: