43
9.4.2
Summary of Clock configuration
functions.
This section summarises all the Clock configuration
functions. When read in conjunction with Fig 21 it
provides a quick aid for configuring the Clock. If
more detail is required, each section of this summary
contains a reference to a full description of the
function.
Display
Summary of function
FunCtion
Instrument function
Defines the function of the instrument.
May be set to:
ELAP5E
Timer
CloC
Clock
All the entries in this Clock
configuration summary assume that
the BA377E-SS is configured as a
Clock by selecting
CLoC
.
See section 9.4.3
di5PLAY
Display format
Defines the clock display format, four
alternatives are available.
Select:
12 : 00 : 00
Twelve hours with seconds
12 : 00
Twelve hours without seconds
24 : 00 : 00
Twenty four hours with seconds
24 : 00
Twenty four hours without seconds
See section 9.4.4
5Et
Set clock display time
Enables the clock displayed time to be
adjusted, the clock resumes operation
when the instrument is returned to the
display mode.
See section 9.4.5
5YnC t
Synchronising time
Defines the time to which the clock
display is set when the reset terminals
RS1 and RS2 are connected together.
The Clock restarts from the
synchronising time when terminals
RS1 and RS2 are disconnected.
See section 9.4.6
Display
Summary of function
oP1
Control output 1 (Optional)
Control output 1 can turn
on
(output
closed) and
off
(output open) twice in
each twenty-four hour period. This
function contains five sub-function:
EnbL
Enables output 1 and defines if
it turns on and off once or twice
in each 24 hours.
oP 1 : on 1
Time when oP 1 turns
on
1
st
time
oP 1 : oFF 1
Time when oP 1 turns
off
1
st
time
oP 1 : on 2
Time when oP 1 turns
on
2
nd
time
oP 1 : oFF 2
Time when oP 1 turns
off
2
nd
time
See section 9.4.7 to 9.4.9
oP2
Control output 2 (Optional)
As control output 1 described above.
See section 9.4.10 to 9.4.12
AC5P
Access control output alarm-times
from display mode.
Contains two sub-functions,
EnbL
which
when activated allows the control output
times to be adjusted from the display
mode (when BA377£-SS is displaying
time). The second sub-function
ACCd
defines a four digit alphanumeric
access code which may be used to
protect access to the alarm-times from
the display mode. Default code
0000
disables this security feature and allows
unrestricted access.
See section 9.4.13
CodE
Security code
Defines a four digit alphanumeric code
that may be used to protect access to
the Clock configuration menu. Default
code
0000
disables this security function
and allows unrestricted access to all
configuration functions when the
(
and
)
buttons are operated simultaneously
in the display mode.
See section 9.4.14
r5Et dEF
Reset to factory defaults
Resets the BA377E-SS to the clock
factory default configuration shown in
section 9.4 when
YE5
is selected and
confirmed by entering
5uRE
.
See section 9.4.15
Summary of Contents for BA377E-SS
Page 1: ...Issue 6 4th May 2021 BA377E SS Rugged one input intrinsically safe Timer or Clock Issue 6...
Page 30: ...30...
Page 31: ...31...
Page 42: ...42...
Page 54: ...54...
Page 55: ...55...
Page 56: ...56...
Page 57: ...57...
Page 58: ...58...
Page 59: ...59...
Page 60: ...60...
Page 61: ...61...
Page 62: ...62...
Page 63: ...63...
Page 64: ...64...
Page 65: ...65...