The second digit should then be
increased until acceptable rate display
stability is once again achieved.
To adjust the filter parameters select
FiLtEr
from the main menu and press
(
to reveal the current setting. The first
digit will be flashing and may be adjusted
using the
&
or
*
button. The
(
button will transfer control to the second
digit. When both are set as required,
return to the
FiLtEr
prompt in the
configuration menu by pressing
)
.
Note:
While adjusting the filter, the rate is
shown on the lower display so that
stability can be assessed.
Step 11 Define the security code
Defining an access security code
prevents unauthorised access to the
configuration menu. Select
CodE
in the
configuration menu and press
(
which
will reveal the existing security code with
the first digit flashing. Enter the new
code 1209 using the
&
or
*
button to
adjust the flashing digit and the
(
button to transfer control to the next
digit. When the new code has been
entered, press
)
to return to the
configuration menu.
Step 12 Return to the display mode
Following completion of configuration,
return the BA367NE to the display mode
by pressing
)
. The instrument will
display
dAtA
followed by
5AVE
while the
configuration changes are stored in
permanent memory.
The BA367NE was assumed to initially have factory
default configuration, therefore the counting edge,
counting direction, local total and local grand total
resets were not reconfigured as they already
complied with the requirements for this example.
During commissioning the debounce and filter
functions may need adjustment to obtain a stable
display.