34
870MHz 28 V/m PM 18 Hz
930MHz 28 V/m PM 18 Hz
1720MHz 28 V/m PM 217
Hz
1845MHz 28 V/m PM 217
Hz
1970MHz 28 V/m PM 217
Hz
2450MHz 28 V/m PM 217
Hz
5240MHz 9 V/m PM 217
Hz
5500MHz 9 V/m PM 217
Hz
5785MHz 9 V/m PM 217
Hz
Conducted Immunity:
IEC 61000-4-6:2015 &
EN61000-4-6:2014
6V rms ISM and amateur
radio bands
Met or exceeded
Voltage Dips and
Interrupts:
IEC 61000-4-11:2005 &
EN61000-4-11:2004
240VAC 500mS (50/60Hz)
70% @ 0°
240VAC 20mS (50/60Hz)
100% @ 0°
240VAC 10mS (50/60Hz)
100%
240VAC 5S (50/60Hz)
100%
100VAC 500mS (50/60Hz)
70% @ 0°
100VAC 20mS (50/60Hz)
100% @ 0°
100VAC 10mS (50/60Hz)
100% @ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
100VAC 5S (50/60Hz)
100%
Met or exceeded
Power Frequency
Magnetic Field:
IEC 61000-4-8:2012 &
EN61000-4-8:2010
30 A/m
(50Hz and 60Hz)
Met or exceeded
Fast Transient Burst:
IEC 61000-4-4:2016 &
EN61000-4-4:2012
± 2kV
Met or exceeded
Surges:
IEC 61000-4-5:2008 &
EN61000-4-5:2006
± 0.5, 1.0 kV L
-L
± 0.5, 1.0, 2.0 kV L
-E 20s
Met or exceeded
ESD:
IEC 61000-4-2:2012 &
EN61000-4-2:2009
+/- 8kV contact
+/- 2, 4, 8,15kV air
Met or exceeded