38
Cleaning
1.
The Gastro
+
™ Gastrolyzer® is moulded with SteriTouch® technology for optimum
infection control and has proven bacterial protection efficacy. However Bedfont®
recommend wiping the instrument and D-piece™ external surfaces with a product
specifically developed for this purpose. Bedfont® provides instrument cleaning wipes.
The D-piece™ cannot be sterilised.
2.
NEVER use alcohol or cleaning agents containing alcohol or other organic solvents as
long term exposure to these vapours will damage the H
2
sensor inside.
3.
Under no circumstances should the instrument be immersed in liquid or splashed with
liquid.
Calibration
The Gastro
+
™ Gastrolyzer® is calibrated at 21°C (±4°C)
before leaving Bedfont®. The instrument
should be calibrated at 21°C (±4°C) as this is the temperature at which we recommend it is to be
used.
The Gastro
+
™ Gastrolyzer® should be calibrated every three months and will give a reminder
during start-up when calibration is due. The calibration gas required is 100ppm hydrogen in air.
Turn on the monitor by pressing the power button once.
Press the settings icon.