6
Remove the bowl liner and dispose. Install a new bowl liner.
7
Replace the black gasket around the chamber opening.
8
Inspect rotor for cracks or damage.
9
Reinstall the rotor. For more information, refer to
10
Plug the power cord into the electrical outlet.
Disinfecting the Rotor
1
Open the centrifuge cover.
2
Unplug the power cord of the external power supply from the electrical outlet.
3
Remove all the tubes and inserts.
4
Remove the rotor. For more information, refer to
Removing and Installing the Rotor
.
5
The rotor and inserts can be soaked in detergent and warm water or a 10% bleach
solution.
6
Do not attempt to clean an insert that has a broken tube. Dispose of the insert
immediately.
7
Dry the rotor and inserts with a clean absorbent paper towel or allow to air dry.
8
Inspect rotor for cracks or damage.
9
Reinstall the rotor. For more information, refer to
10
Plug the power cord of the external power supply into the electrical outlet.
Inspecting the Rotor Speed
The rated speeds can be inspected with a stroboscope or photoelectric tachometer. Point
the tachometer through the transparent cover at the reflective patch located near rotor
center. Do not defeat any safety interlocks while performing the test. If the StatSpin Express
2 fails to achieve an operating speed of 8,500 rpm ±5% in the 120s and 180s settings,
contact Beckman Coulter Customer Support.
Maintenance
Inspecting the Rotor Speed
4-2
55-003910-001FD
Summary of Contents for StatSpin Express 2 M501-22
Page 4: ...Revision History iv 55 003910 001FD ...
Page 16: ...Contents xvi 55 003910 001FD ...
Page 22: ...System Overview Accessories 2 4 55 003910 001FD ...
Page 26: ...Operating Instructions Spinning the Sample 3 4 55 003910 001FD ...
Page 32: ...Specifications Specifications A 2 55 003910 001FD ...
Page 34: ...References References B 2 55 003910 001FD ...
Page 35: ......
Page 36: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved ...