Do not use mobile or cordless telephones and transceivers in the same room as the
system.
•
Do not use medical equipment that can be susceptible to malfunctions caused by Electric
Magnetic Field (EMF) near the system.
Caution
Disconnect the power cord of the external power supply from the electrical outlet
before performing maintenance or inspection.
Caution
Do not spray cleaning solutions directly onto the centrifuge bowl or housing.
Overspray can reach the motor bearings or internal circuitry, causing harm to the
electronics. Before using any cleaning or decontamination methods other than those
recommended by the manufacturer, users should check with the manufacturer that
the proposed method will not damage the equipment.
Caution
During operation maintain a 30 cm (12 inch) clearance around the centrifuge. The
clearance must be free from obstruction and away from the edge of the surface that
the centrifuge is on.
Caution
Follow Universal Precautions with all biological specimens, regardless of whether the
specimen is known to contain an infectious agent. (See
Caution
Use the same size and style tubes in opposite positions. Balance liquid in tubes to
within 0.5mL. If glass tubes are used to balance rotor load, replace the glass balance
tube every ten cycles to prevent possible breakage caused by the repeated stress of
centrifugation. Replace the tube inserts immediately if a tube breaks during
centrifugation.
Caution
Inspect rotor on a routine basis. Rotor lifespan depends on usage. Inspect rotors for
cracks and replace the rotors immediately when any crack or visible wear occurs.
Caution
Never operate the centrifuge without the rotor properly mounted and the locking nut
tightened. Failure to install and secure the rotor correctly can damage the centrifuge.
Safety Notice
Warnings and Cautions
viii
55-003910-001FD
Summary of Contents for StatSpin Express 2 M501-22
Page 4: ...Revision History iv 55 003910 001FD ...
Page 16: ...Contents xvi 55 003910 001FD ...
Page 22: ...System Overview Accessories 2 4 55 003910 001FD ...
Page 26: ...Operating Instructions Spinning the Sample 3 4 55 003910 001FD ...
Page 32: ...Specifications Specifications A 2 55 003910 001FD ...
Page 34: ...References References B 2 55 003910 001FD ...
Page 35: ......
Page 36: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved ...