1
Be sure that the correct tube inserts are installed. The Express 2 centrifuge is provided
with four tube inserts for 13 x 75 mm tubes (Product No. SV01). Other sizes are
available. For more information, refer to
2
Two or four tubes can be processed simultaneously. The rotor must be correctly
balanced for smooth operation.
3
Use the same size and style tubes in opposite positions. Balance liquid in tubes to within
0.5 mL.
Caution
Use the same size and style tubes in opposite positions. Balance liquid in tubes to
within 0.5mL. If glass tubes are used to balance rotor load, replace the glass
balance tube every ten cycles to prevent possible breakage caused by the
repeated stress of centrifugation. Replace the tube inserts immediately if a tube
breaks during centrifugation.
4
To close the lid, firmly apply pressure on the cover directly above the latch until the
gasket compresses completely and the latch engages with the cover.
5
Select the cycle by pressing the cycle selector until the desired LED is illuminated. For
more information, refer to
6
Press the
Start
button.
Unloading
1
When the cycle completes, the rotor decelerates to a complete stop in 25 seconds and
the latch interlock automatically unlocks.
2
Squeeze the black latch pieces together to open cover.
3
Remove tubes carefully to not resuspend the sediment.
Operating Instructions
Spinning the Sample
3
55-003910-001FD
3-3
Summary of Contents for StatSpin Express 2 M501-22
Page 4: ...Revision History iv 55 003910 001FD ...
Page 16: ...Contents xvi 55 003910 001FD ...
Page 22: ...System Overview Accessories 2 4 55 003910 001FD ...
Page 26: ...Operating Instructions Spinning the Sample 3 4 55 003910 001FD ...
Page 32: ...Specifications Specifications A 2 55 003910 001FD ...
Page 34: ...References References B 2 55 003910 001FD ...
Page 35: ......
Page 36: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved ...