![Beckman Coulter Microfuge 20 Series Instructions For Use Manual Download Page 66](http://html1.mh-extra.com/html/beckman-coulter/microfuge-20-series/microfuge-20-series_instructions-for-use-manual_2719742066.webp)
B30162AA
A-4
Installation
Electrical Requirements
Figure A.2
Dimensions of the Refrigerated Centrifuge
WARNING
To reduce the risk of electrical shock, this equipment uses a three-wire electrical
cord and plug to connect the centrifuge to earth-ground. To preserve this safety
feature:
•
Make sure that the matching wall outlet receptacle is properly wired and
earth-grounded. Check that the line voltage agrees with the voltage listed on
the name rating plate affixed to the centrifuge.
•
Never use a three-to-two wire plug adapter.
•
Never use a two-wire extension cord or a two-wire non-grounding type of
multiple-outlet receptacle strip.