B30162AA
2-15
Operation
Summary of Run Procedures
2
5
Enter run parameters:
•
Set run speed
—
MENU
until RPM or RCF displays, use
or
until correct speed displays.
•
Set run duration
—
MENU
until time displays, use
or
until correct time displays.
•
Set run temperature (Microfuge 20R only)
—
MENU
until temperature displays, use
or
until correct temperature displays.
•
Set acceleration/deceleration
—
MENU
until the
Soft-stop/start on, soft-stop on, or soft off
displays, use
or
until correct deceleration displays.
6
Check that all parameters are correct and the door is shut and latched, then press the
START/STOP
key.
WARNING
Never attempt to override the door interlock system while the rotor is spinning.
CAUTION
Do not lift or move the centrifuge while the rotor is spinning.
7
Wait for the set time to count down to zero, or end the run by pressing or holding the
START/STOP
key.
8
After the rotor stops spinning
, press the
OPEN
key to release the door latch; open the door.
9
Unload the rotor according to instructions in the applicable rotor manual.
CAUTION
If disassembly reveals evidence of leakage, you should assume that some fluid
escaped the rotor. Apply appropriate decontamination procedures applied to the
centrifuge and accessories.
Summary of Contents for Microfuge 20 Series
Page 4: ...B30162AA iv Revision Status ...
Page 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Page 14: ...xiv Tables ...
Page 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Page 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Page 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Page 68: ...B30162AA A 6 Installation Test Run ...
Page 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...