![Beckman Coulter Microfuge 20 Series Instructions For Use Manual Download Page 30](http://html1.mh-extra.com/html/beckman-coulter/microfuge-20-series/microfuge-20-series_instructions-for-use-manual_2719742030.webp)
B30162AA
1-10
Description
Available Rotors
Available Rotors
See the applicable rotor manual for information on rotor use, care and maintenance, and rotor
accessories.
Rotor Profile
a
Description
Max
RCF
(
´
g
)
Max
RPM
Max
Capacity
(mL)
Rotor Part
Number
Rotor
Manual
Number
FA241.5
Fixed Angle, 44° Angle
r
max
= 82.0 mm
Rotor code: 2
20,627
15,000
24
´
2.0
B30155
B30178
FA241.5P
Fixed Angle, 43° Angle
r
max
= 82.0 mm
Rotor code: 3
20,627
15,000
24
´
2.0
B30156
B30178
FA4
´
8.2P
Fixed Angle, 44° Angle
r
max
= 66.0 mm
Tube 1 & 8 at
r
max
= 66.0 mm
Tube 2 & 7 at
r
max
= 63.0 mm
Tube 3 & 6 at
r
max
= 60.0 mm
Tube 4 & 5 at
r
max
= 59.0 mm
Rotor code: 4
16,602
15,847
15,093
14,841
15,000
32
´
200
m
L
B31059
B33057
FA361.5
b
Fixed Angle
Inner row:
50° Angle
Outer row:
30° Angle
Inner row:
r
max
= 80 mm
Outer row:
r
max
= 80 mm
Rotor code: 5
20,124
15,000
36
´
2.0
B31057
B33026
a. Refer to the individual manuals for each rotor for more detailed specifications.
b. Rotor is certified for biocontainment by a third-party test facility CAMR, Porton Down, UK, or USARMIID, Ft. Detrick, MD, U.S.A.). Improper use or
maintenance may affect seal integrity and thus containment.
Summary of Contents for Microfuge 20 Series
Page 4: ...B30162AA iv Revision Status ...
Page 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Page 14: ...xiv Tables ...
Page 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Page 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Page 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Page 68: ...B30162AA A 6 Installation Test Run ...
Page 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...