B30162AA
2-4
Operation
Run Procedure
Entering Run Parameters
When the power is applied for the initial use (no previous runs), default values are displayed
(see
). After the initial use, the parameters of the latest previous run are displayed when
power is applied.
When run parameters for a rotor have been entered, as described below, they are retained in the
centrifuge memory. The recalled program can then be used for the current run or altered as
required. See
.
Figure 2.2
Default Parameters on First-Time Use, Microfuge 20R shown
Menu Key Sequence
Pressing the MENU key starts the parameter change sequence. Repeatedly pressing the MENU key
steps through parameter sequence. When the parameter units (rpm, rcf, m & s, soft-off,
soft stop/start on, soft stop on) flash, press
or
to change that parameter. The entire parameter
change sequence follows these steps:
1.
Speed
a.
in RPM
b.
in RCF
2.
Time in minutes and seconds
3.
Temperature (for Microfuge 20R only)
4.
Acceleration/Deceleration Options
a.
Soft off
b.
Soft-stop/start on
c.
Soft-stop on
1.
15,000 rpm
2.
2 min duration
3.
temperature: 20°C
4.
no soft acceleration
Summary of Contents for Microfuge 20 Series
Page 4: ...B30162AA iv Revision Status ...
Page 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Page 14: ...xiv Tables ...
Page 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Page 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Page 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Page 68: ...B30162AA A 6 Installation Test Run ...
Page 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...