B30162AA
2-11
Operation
Run Procedure
2
—
Throughout the run, checks are made to ensure that the rotor does not exceed
set speed.
•
The
SPEED
display indicates the actual rotor speed in RPMs.
—
The RCF can be displayed by pressing he
or
cursor key.
—
This display also shows the time remaining in the run (or
¥
and elapsed time for
continuous operation).
•
The
TEMP
display indicates the actual temperature of the rotor.
WARNING
Do not attempt to override the door interlock system while the rotor is spinning.
CAUTION
Do not lift or move the centrifuge while the rotor is spinning. Do not place items
on the centrifuge during operation.
Short Run (Quick Run) Function
During a short run, the centrifuge accelerates at a maximum rate until the set speed is reached.
The short run function, accessed by pressing the
QUICK RUN
key, is used for short-duration runs
as follows:
NOTE
The Microfuge 20R Refrigerated centrifuge does not have a dedicated
QUICK RUN
button,
but pressing and holding the
START/STOP
button serves the same function.
•
When the
QUICK RUN
key is pressed, the rotor accelerates at a maximum rate to the set speed
and continues to spin as long as long as the
QUICK RUN
key is pressed. (The current run time,
acceleration, and deceleration settings are overridden by the short run function.) When the
QUICK RUN
key is released, the rotor begins decelerating to 0 RPM using maximum deceleration.
•
When the
QUICK RUN
key is pressed, the
TIME
display begins displaying the elapsed seconds.
When the
QUICK RUN
key is released, the seconds stop accumulating.
•
The centrifuge memory retains the parameters of the last run performed before the
QUICK RUN
key was pressed. At the end of a quick run, after the centrifuge door is opened and closed the
previous run parameters are displayed.
To perform a quick run on the non-refrigerated Microfuge 20
1
Close the centrifuge door and push firmly down on both sides of the door front until the
clicking (latching) sound stops.
2
For the Microfuge 20, press and hold the
QUICK RUN
key.
The message
Short Run
and the duration of the quick run is displayed.
Summary of Contents for Microfuge 20 Series
Page 4: ...B30162AA iv Revision Status ...
Page 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Page 14: ...xiv Tables ...
Page 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Page 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Page 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Page 68: ...B30162AA A 6 Installation Test Run ...
Page 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...