B30162AA
4-4
Care and Maintenance
Decontamination
Decontamination
If the centrifuge and/or accessories are contaminated with radioactive or pathogenic solutions,
perform appropriate decontamination procedures. Refer to
Chemical Resistances
(IN-175) to be sure
the decontamination method will not damage any part of the centrifuge.
Sterilization and Disinfection
The centrifuge is finished with urethane paint. Ethanol (70%) may be used on this surface.
See
Chemical Resistances
for more information regarding chemical resistance of centrifuge and
accessory materials.
CAUTION
Ethanol is a flammability hazard. Do no use it in or near operating centrifuges.
While Beckman Coulter has tested these methods and found that they do not damage the
centrifuge, no guarantee of sterility or disinfection is expressed or implied. When sterilization or
disinfection is a concern, consult your laboratory safety officer regarding proper methods to use.
Storage and Transport
Storage
Before storing a centrifuge for an extended period, return it to its original shipping container
to protect it from dust and dirt. Reinsert the shipping foam (removed at time of centrifuge
installation) into the chamber, making sure the drive shaft is stabilized in the hole in the foam.
Temperature and humidity conditions for storage should meet the environmental requirements
described in the specifications given in
.
The centrifuge can be stored in its original packaging for up to one year.
•
Store the centrifuge only in dry rooms.
•
The storage temperature must be above –20°C.
•
If the centrifuge needs to be stored for more than one year or needs to be shipped overseas,
contact Beckman Coulter Customer Service.
Notes on Transport
•
When lifting the centrifuge always reach under the centrifuge from the side.
•
Use suitable packaging and, if possible, use the original packaging.
•
Retain the packaging for possible future transport of the centrifuge.
Summary of Contents for Microfuge 20 Series
Page 4: ...B30162AA iv Revision Status ...
Page 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Page 14: ...xiv Tables ...
Page 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Page 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Page 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Page 68: ...B30162AA A 6 Installation Test Run ...
Page 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...