Section III: Sequencing Troubleshooting Quick Reference Table
28
CEQ™ 8000 Series Genetic Analysis System
Low Raw Signal and
Erratic or Crashed
current profiles
Preheat treatment on plasmid
samples was not performed
(plasmid/cosmids/YAC/BAC/
PAC samples).
Preheat all plasmids templates
The preheat treatment was
carried out with all the
sequencing reagents added.
Preheat ONLY the DNA and WATER
The cycling program had an
extra step in the beginning for
the preheat treatment.
Do not
add an extra step in the
beginning of the cycling program
Too much DNA loaded onto
capillaries.
Reduce the amount of DNA—follow
protocol recommendations
Sample contains
contaminating ions that are
injected onto capillaries- (ions
from contaminating salts, bad
formamide etc.).
Check method of DNA purification
Incorrect method of template
purification e.g.- CsCl,
Phenol/Chloroform, or
insufficient removal of
Ethanol during template
preparation etc.
Refer to the CEQ 2000 DTCS
chemistry Protocol, 718119-AB, pg.
18 and Application Note A-1872A for
detail on the preheat treatment
procedure.
Use supplied SLS for sample
resuspension
No signal associated
with normal
Current Profile
(No signal even from
the Unincorporated
dyes)
Lost Pellet
Use “gel loading pipette tips” to
remove ethanol during precipitation
Tips of capillary broken
Replace capillary array if tips are
broken
ddNTPs not added
Check that all reagents were added
Sample not resuspended in
formamide
Only use SLS for sample resuspension
Not enough sample in well
(though in most cases will see
a current problem)
Use at least 30µl of SLS to resuspend
the samples
PROBLEM
RAW DATA
SIGNAL /
CURRENT
PROFILE
POTENTIAL SOURCE OF
PROBLEM
POSSIBLE SOLUTIONS