Section III: Sequencing Troubleshooting Quick Reference Table
Sequence Analysis Troubleshooting Guide
27
Section III: Sequencing Troubleshooting Quick
Reference Table
PROBLEM
RAW DATA
SIGNAL /
CURRENT
PROFILE
POTENTIAL SOURCE OF
PROBLEM
POSSIBLE SOLUTIONS
• Irreproducibility
• Short Read lengths
• Inaccurate base
calling
• Data not analyzed
Low Raw Signal and
Normal current
profile
Not enough DNA
Increase the [DNA], always quantify
the DNA
Poor Thermal Cycling
Check Thermal Cycler
Problem with Thermal Cycler
Check that the correct method was
chosen on the thermal Cycler
Bad Formamide
Use supplied SLS for sample
resuspension
Incorrect primer concentration Check Primer Concentration
Incomplete resuspension of
pellet in formamide
Make sure that the pellet is properly
resuspended in formamide
or
If the pUC18 Control Template does
not meet the specifications
Problem in post reaction
cleanup of samples
Problem with Ethanol precipitation
Check that glycogen was added
Centrifugation at 4
°
C
pH of Sodium Acetate
Concentration of EDTA
Incorrect mineral oil
Check that supplied mineral oil was
added to samples in plate
No mineral added
Check that mineral oil was added to
samples in plate