Section II: Troubleshooting a CEQ System Problem
20
CEQ™ 8000 Series Genetic Analysis System
Any one of the previously mentioned problems will yield low Raw Data signal similar
to that shown in Figure 14.
Corrective Actions
1.
Check thermal cycler for correct performance. If you are using a robocycler
optimization will be required.
2.
Make sure the temperature ramp is set to maximum.
3.
Check primer melting temperature, if the annealing temperature is too high for
the primer you will need to lower it.
4.
Make sure you are using PCR tubes or plates. The CEQ sample plate has been
proven to work on the MJ PTC-200 and P/E 9600 thermal cyclers.
5.
Make sure strip caps, A sealing film, or sealing mat are put on the PCR plate
correctly. The lack of a tight seal will lead to evaporation of water from the tubes
/ plate which will change the buffer concentration in the reaction mixture.
Figure 14: Low Raw Data signal