Section II: Troubleshooting a CEQ System Problem
8
CEQ™ 8000 Series Genetic Analysis System
If the current is unstable or erratic, then the migration times of the bases will be
adversely affected and this can lead to inaccuracies in the base calling. During the time
that the current is not being maintained, as shown in Figure 4, the base migration will
be slowed and the accuracy of base calling will be low. In addition to inaccuracies in
the base call the total read length will also be diminished simply due to the fact that the
migration of the bases has been slowed from the normal run conditions.
In other cases, current failures can be more significant. Figure 5 shows a current
profile that fails to reach the final running current after the normal ramping. Note that
the current never reaches the normal running current of 5–9 µAmps and also declines
rapidly to less than 2 µAmps.
Figure 4:
Erratic Current
Figure 5:
Crashed Current
Current ramps normally
but then dips down and
recovers.
Does not achieve normal current level
Rapid decline in the current
Low separating current