B20063AC
3-6
Preparing Cytomat Devices
Setting Up Optical Tables
5
To level the table, place a level across the optical platforms and adjust the feet by turning them
clockwise or counterclockwise as necessary.
NOTE
Contact the nearest Beckman Coulter Service Representative if assistance is required to set up
the optical table.
Setting Up a .6 m x .9 m Optical Table for the Biomek NX
P
Instrument and the
Microplate Hotel
Setting up an optical table should be done under the direct supervision or with the detailed input
of the Beckman Coulter Field Service Engineer.
To set up the .6 m x .9 m optical table:
1
Screw the white plastic feet into the table legs all the way (see