B20063AC

2-7

Preparing the Biomek NXP Instrument for Setup

Moving the NX

P

 Instrument to the Installation Site  

2

To move the instrument and accessories to the site:

1

With one person on each side of the Biomek NX

P

 instrument, grasp the handhold at the rear of 

the instrument with one hand and grasp the end of the base casting with the other hand.

CAUTION

When lifting the Biomek NX

P

, grasp it only from positions on the base casting. Do 

not grasp the ends of the light curtain that extend across the front of the 

instrument.

2

Lift the instrument straight up using the hand positions provided in the base casting and place 
it onto a roll-around cart.

CAUTION
To prevent hands from being pinched, carefully set the back of the instrument 

down first and then lower the front.

3

Set the back of the instrument down first and then carefully lower the front of the instrument.

4

Inspect the pallet to ensure no parts, such as leveling feet, have worked loose in shipping or 
been left on the pallet.

5

Place the accessories onto a second roll-around cart.

6

Roll both carts to the installation site.

Summary of Contents for Biomek NXP

Page 1: ...B20063AC April 2019 Preinstallation Manual Biomek NXP Laboratory Automation Workstation Beckman Coulter Inc 250 S Kraemer Blvd Brea CA 92821 U S A...

Page 2: ...crosoft Corporation in the United States and or other countries All other trademarks service marks products or services are trademarks or registered trademarks of their respective holders May be cover...

Page 3: ...uirements Table 4 1 Electrical Components and Requirements Related Documents Note Changes that are part of the most recent revision are indicated in text by a bar in the margin of the amended page Thi...

Page 4: ...B20063AC iv Revision History...

Page 5: ...any other automated laboratory analyzer Alerts for Danger Warning Caution Important and Note DANGER DANGER indicates an imminently hazardous situation which if not avoided will result in death or seri...

Page 6: ...ng of equipment requiring the removal of covers or panels High Voltage This symbol indicates the potential of an electrical shock hazard existing from a high voltage source and that all safety instruc...

Page 7: ...njury Keep doors covers and panels closed and secured in place while the instrument is in use Take full advantage of the safety features of the instrument Do not defeat safety interlocks and sensors A...

Page 8: ...ulter Representative WARNING To prevent possible injury wear safety glasses or safety goggles when removing metal strapping Strapping may spring back causing injury WARNING To prevent possible injury...

Page 9: ...Do not grasp the ends of the large cover at the top of the instrument CAUTION The optical table is heavy and is cumbersome To avoid injury two people are needed to unpack the optical table from the s...

Page 10: ...B20063AC x Safety Notice Warnings and Cautions Found in this Manual...

Page 11: ...luid Requirements NXP S8 only 1 8 CHAPTER 2 Preparing the Biomek NXP Instrument for Setup 2 1 Overview 2 1 Moving the NXP Instrument Crate to the Unpacking Location 2 1 Unpacking the NXP Instrument 2...

Page 12: ...tical Table for the Biomek NXP Instrument and the Microplate Hotel 3 6 Optical Table Layout 3 8 Unpacking Cytomat Devices 3 11 CHAPTER 4 Preparing Other Integrated Devices 4 1 Overview 4 1 Dimensions...

Page 13: ...nts for the NXP Instrument Model A31840 2 4 2 3 U Brackets 2 5 3 1 Assembled 1 m x 1 m Optical Table 3 5 3 2 Assembled 6 m x 9 m Optical Table 3 7 3 3 Layout Showing 6 m x 9 m and 1 m x 1 m Optical Ta...

Page 14: ...1 5 1 4 Electrical Components and Requirements 1 6 2 1 Accessories by Model Number 2 4 3 1 Cytomat 2C Incubator Specifications 3 2 3 2 Cytomat 6001 HS Incubator Specifications 3 2 3 3 Cytomat Micropl...

Page 15: ...P instrument efficiently upon arrival Appropriate site preparation includes making sure that System specification requirements are met when the Biomek NXP instrument is installed See Biomek NXP System...

Page 16: ...15 30 C System fluid should be degassed for 24 hours prior to use Dimensions base unit NX MC 36 in L x 31 in W x 34 in H 91 44 cm L x 78 74 cm W x 86 36 cm H NX S8 36 in L x 31 in W x 44 in H 91 44 cm...

Page 17: ...for the instrument to overhang the bench Biomek Controller Windows 7 Biomek Software Version 4 1 The following specifications are the minimum requirements needed Processor Intel Core i5 2400 3 1 GHZ o...

Page 18: ...in the above table do not include space for the automation controller or additional accessories A lab bench with shelving for accessories underneath the instrument is desirable 25 in L x 24 in W 63 5...

Page 19: ...Component Weight Biomek NXP Instrument Approximately 190 lbs 86 2 kg Tip Loader ALP 10 lbs 4 5 kg 1 x 1 Passive ALP 2 lbs 9 kg 1 x 3 Passive ALP 4 lbs 1 8 kg 1 x 5 Passive ALP 6 lbs 2 7 kg 4 x 3 Pass...

Page 20: ...essure requirements Air flow of 2 25 cfm 60 psi 0 064 m3 min 276 kPa NOTE Make sure the lab site has a clean dry air supply available at the lab bench A 1 4 in or 1 8 in NPT female fitting to connect...

Page 21: ...uring operation than shop air Figure 1 1 Recommendation for an Air Pressure Regulation System Configuration Vacuum Supply Requirements If conducting PCR or Nucleic Acid Preparation applications that r...

Page 22: ...ure is 59 86 F 15 30 C NOTE System fluid should be degassed for 24 hours prior to system installation Degassing the system fluid is accomplished by letting the system fluid rest in the supply containe...

Page 23: ...the NXP Instrument Inspecting the NXP Instrument Moving the NXP Instrument to the Installation Site Placing the NXP Instrument on the Lab Bench Contacting Your Field Service Engineer Moving the NXP I...

Page 24: ...NXP instrument unpacking procedure requires two people Required Tools Safety glasses or goggles Steel strapping cutter Socket wrench to fit 1 4 inch hex head bolt Hex bit screwdriver Allen wrench Phi...

Page 25: ...ith caution Strapping may spring back when cut causing injury 1 Wearing safety glasses or goggles carefully cut the steel bands securing the Biomek NXP instrument accessories box and instrument box to...

Page 26: ...e 2 1 Accessories by Model Number A31839 A31840 A31841 Span 8 Tube Tool Kit Span 8 Tube Tool Kit N A Biomek Starter Kit Biomek Starter Kit Biomek Starter Kit Span 8 Tip Wash Kit Span 8 Tip Wash Kit N...

Page 27: ...box to the pallet 2 Remove the piece of plywood from the top of the instrument box and set it aside The instrument box consists of the tube tray cap and shell which cover the Biomek NXP instrument on...

Page 28: ...ument to ensure that all parts have been included with the instrument If you discover any damaged or missing parts contact your Beckman Coulter Service Representative immediately Moving the NXP Instru...

Page 29: ...sp the ends of the light curtain that extend across the front of the instrument 2 Lift the instrument straight up using the hand positions provided in the base casting and place it onto a roll around...

Page 30: ...at the top of the instrument 1 With one person on each side of the Biomek NXP instrument grasp the handhold at the rear of the instrument with one hand and the end of the base casting with the other...

Page 31: ...t physically assist in preparing the site and unpacking the devices NOTE A Cytomat 6001 incubator or microplate hotel must be unpacked and set on the optical tables for the Beckman Coulter Field Servi...

Page 32: ...H 49 1 cm L x 53 cm W x 110 1 cm H Crated Weight 311 lbs 141 kg Device Weight 176 lbs 80 kg Electrical Requirements 100 120 VAC 6 amps 230 VAC 6 amps Table 3 2 Cytomat 6001 HS Incubator Specification...

Page 33: ...be taken to the laboratory unpacked and the table s assembled at the proper workspace area A Cytomat 6001 should be placed on a 1 m x 1 m optical table refer to Setting Up a 1 m x 1 m Optical Table fo...

Page 34: ...and discard the bag covering the table Setting Up a 1 m x 1 m Optical Table for the Cytomat 6001 Setting up an optical table should be done under the direct supervision or with the detailed input of...

Page 35: ...iomek NXP instrument and Cytomat devices see CHAPTER 1 Preparing the Site CHAPTER 2 Preparing the Biomek NXP Instrument for Setup and the section titled Cytomat Device Specifications 3 Tilt the shelf...

Page 36: ...n Coulter Service Representative if assistance is required to set up the optical table Setting Up a 6 m x 9 m Optical Table for the Biomek NXP Instrument and the Microplate Hotel Setting up an optical...

Page 37: ...te CHAPTER 2 Preparing the Biomek NXP Instrument for Setup and the section titled Cytomat Device Specifications 3 Tilt the shelf at an angle and maneuver it between the legs of a table Set it down acr...

Page 38: ...ets is determined by the system layout see Figure 3 3 NOTE Contact the nearest Beckman Coulter Service Representative if assistance is required to set up the optical table Optical Table Layout The opt...

Page 39: ...B20063AC 3 9 Preparing Cytomat Devices Setting Up Optical Tables 3 Figure 3 3 Layout Showing 6 m x 9 m and 1 m x 1 m Optical Tables...

Page 40: ...B20063AC 3 10 Preparing Cytomat Devices Setting Up Optical Tables Figure 3 4 Layout for 6 m x 9 m Optical Table...

Page 41: ...t 2C incubator is supplied with a custom sized table that is set up by the field service engineer 1 Remove the top panel of the crate with a Phillips head screwdriver and set it aside WARNING To preve...

Page 42: ...B20063AC 3 12 Preparing Cytomat Devices Unpacking Cytomat Devices...

Page 43: ...grated Devices electrical supply and power are adequate for operation see Electrical Component Requirements Dimensions for Integrated Devices Use Figure 4 1 and Figure 4 2 to determine adequate space...

Page 44: ...everal Integrated Devices 1 Cytomat Microplate Hotel 2 600mm x 900mm Optical Table 3 Standard Cytomat ALP with bar code reader in Cytomat ALP left position for MC See note 4 in Figure 4 1 4 600mm x 90...

Page 45: ...and Requirements Electrical Component Electrical Requirements Bar Code Reader 110 220 VAC 1 amp max Biomek NXP instrument 100 240 VAC 4 amps max Circulating Reservoir Tip Box ALP 115 VAC 1 5 amps 220...

Page 46: ...B20063AC 4 4 Preparing Other Integrated Devices Electrical Component Requirements...

Page 47: ...els 1 1 Site Preparation 1 1 Preparing Cytomat Devices 3 1 Cytomat Device Specifications 3 2 Estimating Required Workspace Dimensions 3 3 Setting Up Optical Tables 3 3 Unpacking Cytomat Devices 3 11 P...

Page 48: ...luid Requirements 1 8 System Specifications 1 2 U Unpacking an Optical Table 3 3 Unpacking Cytomat Devices 3 11 Unpacking the Accessories 2 3 Unpacking the Biomek NXP Instrument 2 2 Required Tools 2 2...

Page 49: ...oactive hazardous material 4 The Company shall be released from all obligations under all warranties either expressed or implied if any product covered hereby is repaired or modified by persons other...

Page 50: ...eseeability or the form of action whether in contract tort including negligence breach of warranty strict liability or otherwise and including but not limited to damages resulting from loss of data lo...

Page 51: ......

Page 52: ...ide for Biomek 3 3 Methods PN B32462 Biomek NX and NXP Laboratory Automation Workstations Multichannel Pod User s Manual PN 987892 Biomek NX and NXP Laboratory Automation Workstations Span 8 Pod Hardw...

Reviews: