5-4
Maintenance
While Beckman Coulter has tested ethanol (70%) and found that it
does not damage the centrifuge, no guarantee of sterility or disinfec-
tion is expressed or implied. When sterilization or disinfection is a
concern, consult your laboratory safety officer regarding proper
methods to use.
REPLACING THE AIR FILTER
Check the air filter (Figure 5-1) regularly and replace it about once a
year, or more often if it looks dirty. The air filter is not fastened to the
centrifuge, so no tools are required for removal or installation.
1. To remove the air filter, grasp the top edge and lift the filter
straight up.
2. Install a new filter by holding the top edge and inserting the
bottom edge into the slot in the air deflector strip. Lower the filter
until the bottom edge rests on the brackets.
Figure 5-1. Air Filter Replacement
Air Deflector
Filter Brackets
Filter