![Beckman Coulter Avanti J-E Instruction Manual Download Page 25](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-e/avanti-j-e_instruction-manual_2719708025.webp)
1-11
Description
AVAILABLE ROTORS
Refer to the applicable rotor manual for complete rotor descriptions.
a
Maximum speeds are based on a solution density of 1.2 g/mL in all rotors.
b
Maximum speed for the rotor in the Avanti
®
J-E centrifuge.
c
At 35°C ambient temperature and 95 percent humidity.
d
Maximum speed in the Avanti J-E for the rotor with magnets; without magnets maximum is 14 000 rpm. (Maximum speed at 2°C in
a 50-Hz centrifuge is 14 000 rpm.)
e
Maximum speed for rotor without magnets is 13 000 rpm.
f
Temperature performance for the F14BCI-14x50cy rotor in the Avanti J-E is as follows: 12°C minimum at 14 000 rpm (at 35°C ambient);
4°C minimum at 13 000 rpm (35°C ambient).
g
Maximum speed for rotor without magnets is 6 300 rpm.
Rotor Profile and Description
Rotor
Entry
Code
Max
RPM
a
Max RCF
(
×
g
)
Max
Capacity
Rotor
Manual
Number
JA-25.50 Fixed Angle, 34° (8 place)
r
max
= 108 mm
25.5
21 000
b
(18 000
@ 2°C)
c
53 300
8
×
50 mL
J-TB-056
JA-21 Fixed Angle, 40° (18 place)
r
max
= 102 mm
21
21 000
(18 000
@ 2°C)
c
50 400
18
×
10 mL
J-TB-002
JA-20.1 Fixed Angle, 23° (32 place)
r
max
= 115 mm
(outer row)
r
max
= 98 mm
(inner row)
20.1
20 000
(17 000
@ 2°C)
c
51 500
43 900
32
×
15 mL
J-TB-022
JA-20 Fixed Angle, 34° (8 place)
r
max
= 108 mm
20
20 000
(17 000
@ 2°C)
c
48 400
8
×
50 mL
J-TB-003
JA-18 Fixed Angle, 23° (10 place)
r
max
= 132 mm
18
16 000
d
(14 000
@ 2°C)
c
37 800
10
×
100 mL
J-TB-035
JA-17 Fixed Angle, 25° (6 place)
r
max
= 132 mm
17
17 000
e
(15 000
@ 2°C)
c
39 800
14
×
50 mL
J-TB-017