![Beckman Coulter Avanti J-E Instruction Manual Download Page 36](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-e/avanti-j-e_instruction-manual_2719708036.webp)
3-2
Operation
!
WARNING
Do not use the centrifuge in the vicinity of
flammable liquids or vapors, and do not run
such materials in the instrument. Do not lean
on the instrument or place items on it while it
is operating.
SUMMARY OF RUN PROCEDURES
For runs at other than room temperature, refrigerate or warm
the rotor beforehand for fast equilibration.
PROGRAMMED RUN
1
POWER
Turn the power switch on (
I
).
2
[
DOOR
]
Press [
DOOR
] to unlock the chamber door; lift the door open.
3
Install the rotor according to the applicable rotor manual, then
close the chamber door and press [
DOOR
].
4
[
PROGRAM
]
Press [
PROGRAM
], use the keypad to enter the required program
number, then press [
ENTER
]. Press [
SHOW SET
] to display program
parameters.
5
[
ENTER
]
[
START
]
Check that all parameters are correct and that the door is closed.
Press [
ENTER
], then press [
START
] (within 5 seconds).
6
[
STOP
]
Wait for the run to end, or end the run by pressing [
STOP
].
7
[
DOOR
]
When the rotor stops (a tone sounds), press [
DOOR
] to unlock the
chamber door; lift the door open to remove the rotor.