2-2
Preinstallation Requirements
If it is necessary to move the centrifuge, maintain the following
conditions.
• Select a location away from heat-producing laboratory equipment,
with sufficient ventilation to allow for heat dissipation.
• The centrifuge must have adequate air ventilation to ensure
compliance to local requirements for vapors produced during
centrifuge operation.
• Position the centrifuge on a level floor that can support the weight
of the centrifuge and resist vibration.
• In addition to space for the instrument itself, allow a 7.7-cm (3-in.)
clearance on each side of the instrument and a 16-cm (6.25-in.)
clearance behind the instrument for air circulation. The centrifuge
must have adequate air ventilation to ensure compliance to local
requirements for vapors produced during operation.
Figure 2-1. Dimensions of the Avanti
®
J-E Centrifuge
34.0
86.4
25.0
63.5
3.0
7.7
3.0
7.7
31.5
80.0
6.25
16.0
60.0
152.4
cm
in.