![Beckman Coulter Avanti J-E Instruction Manual Download Page 27](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-e/avanti-j-e_instruction-manual_2719708027.webp)
1-13
Description
a
Maximum speeds are based on a solution density of 1.2 g/mL in all rotors.
b
Maximum speed for the rotor in the Avanti
®
J-E centrifuge.
c
At 35°C ambient temperature and 95 percent humidity.
d
Maximum speed in the Avanti J-E for the rotor with magnets; without magnets maximum is 14 000 rpm. (Maximum speed at 2°C in
a 50-Hz centrifuge is 14 000 rpm.)
e
Maximum speed for rotor without magnets is 13 000 rpm.
f
Temperature performance for the F14BCI-14x50cy rotor in the Avanti J-E is as follows: 12°C minimum at 14 000 rpm (at 35°C ambient);
4°C minimum at 13 000 rpm (35°C ambient).
g
Maximum speed for rotor without magnets is 6 300 rpm.
Rotor Profile and Description
Rotor
Entry
Code
Max
RPM
a
Max
RCF
(
×
g
)
Max
Capacity
Rotor
Manual
Number
JLA-10.500 Fixed Angle, 20° (6 place)
r
max
= 166 mm
10.5
10 000
g
18 600
6
×
500 mL
J-TB-048
F10BCI-6x500y Fixed Angle, 23°
(6 place)
r
max
= 158 mm
500
10 000
17 696
6
×
500 mL
—
JLA-9.1000 Fixed Angle, 20° (4 place)
r
max
= 185 mm
9.1
6 300
b
8 230
4
×
1000 mL
J-TB-073
JS-13.1 Swinging Bucket (6 place)
r
max
= 140 mm
13
13 000
(12 000
@
2°C)
c
26 500
6
×
50 mL
J-TB-036
JS-5.3 Swinging Bucket (4 place)
r
max
= 194.8 mm
5.3
5 300
6 130
4
×
500 mL
24 microplates
8 deep-well plates
4 square-well plates
J-TB-089