MPS 500 – User Manual Rev. 2.1
BeamTec GmbH
Wolfgang-Paul-Str. 4, 89081 Ulm, Germany
http:
, e-mail: [email protected] tel.: +49 731 146620-0
34
Fig.4.2.5 Constant Voltage mode set
NOTICE: If MPS500 is in standby mode, the present voltage value is the target value
which will be reached after switching to the operating mode -> HV ON.
Voltage value will follow according to the high voltage ramp which is preset in Ramp
menu.
Also if any other parameters ( power or current ) exceed max preset value ( see chap.
4.3.1, Fig. 4.3.2 ), the voltage regulation mode will be limited so as not to exceed this
value.
IMPORTANT
:
The parameters will be stored automatically in the memory whenever
they are changed.
4.2.5.
C
ONSTANT
C
URRENT
M
ODE
S
ET
Use arrow buttons
▲
or
▼
to set the pointer cursor on to Current value – Fig.4.2.6
Turn rotary encoder to increase /decrease the value.
NOTICE: current value set is limited by maximum value which is preset and stored in
the internal memory of the device. Imax value can be changed from within the Limits
menu – see chap. 4.3.1
Fig.4.2.6 Constant Current mode set
NOTICE: If MPS500 is in standby mode, present current value is the target value
which will be reached after switching to the operating mode -> HV ON.
Current value will follow according to the high voltage ramp which is preset in Ramp
menu.
Also if any other parameters ( power or voltage ) exceed max preset value ( see chap.
4.3.1, Fig. 4.3.2 ), the current regulation mode will be limited so as not to exceed this
value.
U: 650V
<
I: 0mA
P: 70W
T:15:00
U: 650V I: 50mA
<
P: 70W
T:15:00