21
interruptions
and voltage
variations on
power supply
input lines
iEc 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not
applicable
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the TH30F
requires continued
operation during power
mains interruptions, it is
recommended that the
TH30F be powered from
an uninterruptible power
supply or a battery.
power
frequency
(50/60 hz)
magnetic field
iEc 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE : UT is the a.c. mains voltage prior to application of the test level.
rEcoMMEnDED sEpArAtion DistAncEs bEtWEEn portAbLE AnD MobiLE rF
coMMUnicAtions EqUipMEnt AnD thE ME EqUipMEnt or ME sYstEM
The TH30F is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the TH30F can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the TH30F as recommended
below, according to the maximum output power of the communications equipment.
separation distance according to frequency of transmitter (m)
rated maximum
output power of
transmitter (W)
150 khz
to 80 Mhz
d = 1,2 √p
80 Mhz
to 800 Mhz
d = 1,2 √p
800 Mhz to 2,5 ghz
d = 2,3 √p
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
notE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
notE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
WArrAntY
Manufacture Date: as the serial number( please open the battery cover, it is shown on the
inside of the device.) Ex.SN:E912A000001, the first “E” is External, the second number “9”
is the manufacture year 2009, the third and the fourth numb er “12” is the manufacture
month, the others is the serial number.
Note: The thermometer is calibrated at the time of manufacture. If at any time you question
the accuracy of temperature measurements, please contact the dealers or nearest service
address. This thermometer is guarranteed for 48 months.
The CE mark and Notified Body Registration Numbers, the requirement of Annex II from
Medical Device Directive 93/42/EEC are met
The logo which appears on this product of a crossed out wheelie bin
indicates that the product should not be disposed of in landfill with normal
household waste. In order to avoid damaging the environment or human
life, please separate this product from other waste in order to ensure that
it is recycled in an environmentally safe manner. For further details on
existing collection sites, please contact the local administrative authorities
or the retailer from which you purchased this product.