17
www.bdsensors.de
ASM 400
12. Technical Data
output signal
2-wire-system:
4 … 20 mA
3-wire-system:
0 … 10 V
contact
Number, type
0, 1, or 2 independent PNP outputs
Switching
performance
standard:
contact rating max. 125 mA, short-circuit resistant
Ex-protection: max. switching current
2
: 70 mA;
max. permissible inductivity per contact: 4.7 mH
Repeatability:
≤ ±
0.1 % FSO
Switching frequency:
max. 10 Hz
Switching cycles:
> 100 x 10
6
Delay time:
0 ... 100 s
Electrical protection
Short-circuit protection permanent
Reverse polarity
protection:
no damage, but also no function
Option Ex protection
AX11-ASM 400:
zone (0) 1: II (1) 2 G EEx ia IIC T4 (only with 4 ... 20 mA / 2-wire)
safety technical maximum values: V
i
= 28 V,
Σ
I
i
= 93 mA,
Σ
P
i
= 660 mW
Display
Type
4-digit, red LED display, digit height 7 mm, digit width 4.85 mm
Range:
-1999 ... +9999
Accuracy:
0.1 % ± 1 digit
Digital damping:
0.3 … 30 sec (programmable)
Update value:
0.0 … 10 sec (programmable)
Mechanical stability
Vibration:
5 g RMS (20 ... 2000 Hz)
Shock:
100 g / 11 ms
Permissible temperatures
Electronics /
environment:
-25 ... 85 °C
Storage:
-40 ... 85 °C
Materials
Display housing
PA 6.6, polycarbonate
Miscellaneous
Cords length
150 mm
Weight
approx. 35 g
Data memory:
non-volatile EEPROM
ingress protection
IP 00 (front side: IP 65)
The ingress protection of the manufactured device depends on the
housing, mounted by the customer.
2
the real switching current in the application depends on the power supply unit
Summary of Contents for ASM 400
Page 18: ...ASM 400 Dimensions...
Page 19: ...19 www bdsensors de ASM 400...