background image

 

   

Instructions pH FUN V1

 

 

 

Page 20 of 28 

Performing the pH7 calibration 

 

 

Danger warning:

  

The regular performing of a pH7 calibration is an important prerequisite for a correct measurement 
of the pH value and thus for the correct control and dosing function of the pH FUN. A missing or 
incorrect pH7 calibration can have as a result a pH value in the pool water that is too high or too 
low, which in the extreme case, can lead to a health hazard. 

It is therefore imperative that a pH7 calibration be performed with the initial operation as well as 
subsequently at least every 3 months. A new pH7 calibration must also be performed again after a 
replacement of the pH electrode. 

Moreover, the pH value in the pool water should be regularly tested using pH test strips or a 
Phenol Red colour test (commercially available). 

 

For the pH7 calibration, you will require the pH7 buffer solution, which is included in the scope of supply of the pH FUN. 
 

Please follow exactly the following sequence step-by-step for the pH7 calibration: 

1. 

In the normal view of the pH FUN, press the 

 button (at least 2 seconds) to call up the menu. 

2. 

A screen page for the pH7-Calib. appears immediately: 

 

pH7-Calib. OK_ 

7.09 pH 

                              

 The currently measured pH value is displayed in this page, in the example 7.09 pH. 

 

3. 

Stop the measurement water flow through the measurement cell by closing the appropriate shut-off valve. 
Open the container with the pH7 buffer solution. 

4. 

Remove the plastic cover of the pH FUN measurement cell and carefully unscrew the pH electrode out of the 
measurement cell. Loosen the BNC connector of the electrode on the pH FUN device beforehand so that the 
electrode cable does not get twisted when unscrewing the electrode. Handle the electrode by its plastic shaft 
only. 

5. 

Rinse the electrode in clean water and then dry it carefully using a soft cloth. 

6. 

Now immerse the electrode in the container with the pH7 buffer solution. You should not sway the electrode but 
rather immerse it to the bottom of the container and hold it as still as possible. 

7. 

Observe the displayed pH value on the pH FUN. After immersion in the pH7 buffer solution, it will usually 
change slowly. After approx. 1-2 minutes in the buffer solution, the display should stabilise again and only 
fluctuate by a maximum of 0.01..0.02 pH. 

8. 

When the displayed pH value for the buffer solution has stabilised, press the 

 button to perform the pH7 

calibration. 

9. 

The pH measured value on the pH FUN is corrected to 7.00 pH by the calibration. After the calibration, the 
value 7.00 pH should therefore appear in the display of the pH FUN, whereby slight fluctuations in the range of 
approx.  6.97...7.03 pH are normal and acceptable. 

10.  Now press the 

 button again (at least 2 seconds) to return to the normal view of the pH FUN. 

11.  Remove the pH electrode from the buffer solution, rinse it in clean water and then dry it carefully using a soft 

cloth. 

12.  Screw the electrode back into the measurement cell again. Tighten the electrode firmly by hand. Ensure a 

secure sealing. 

13.  Now connect the BNC connector of the electrode to the pH FUN device again and attach the plastic covering of 

the pH FUN measurement cell again. 

14.  Start the measurement water flow through the measurement cell again by opening the appropriate shut-off 

valve. 

If the pH7 calibration does not function, please observe the information in Chapter 9.5 The pH7 calibration does not 
work. 

Summary of Contents for pH FUN

Page 1: ...Instructions pH FUN V1 Page 1 of 28 Instruction manual pH FUN...

Page 2: ...ng field of the pH FUN 13 4 1 1 Button functions in the normal view 14 4 1 2 Performing a parameter setting 14 4 2 The pH FUN menu 14 4 3 Setting options of the pH FUN 16 4 3 1 Menu navigation 16 4 3...

Page 3: ...d operated by qualified expert personnel General danger warning Changing the system settings default values can be dangerous to persons under certain circumstances Therefore changes must only be made...

Page 4: ...ok after the disinfection The use of BAYROL algaecide and flocculant products is additionally recommended The product program is perfectly matched to guarantee a perfect water quality with very little...

Page 5: ...ck that the float in the measurement chamber can float freely Adjust the water flow through the cell so that the float just barely lies against the upper end in its guide bore Ensure that connections...

Page 6: ...lowed The equipment should be installed by a licensed electrician The supply voltage for the device may not exceed 240V 50 Hz The allowable operating temperature range is from 0 to 50 O C and the allo...

Page 7: ...to the measurement cell it must be fed directly into the circulating line You will require the following accessories available at your BAYROL dealer 100 509 PE hose length depending on the installatio...

Page 8: ...Measurement water intake 6 Heat exchanger 7 Sand filter 8 Measurement water extractor 9 Circulating pump 10 Container for pH Minus liquid 11 pH Minus suction line 1 pH FUN Chlorine 2 Swimming pool 3 M...

Page 9: ...available Use Europe for all countries not listed Confirm with Pool volume Enter the volume of the swimming pool in m cubic meter using the arrow keys and confirm with The input of the pool volume se...

Page 10: ...Performing the pH7 calib Replace the hoses of the dosing pumps 3 1 1 Checking the hoses Check all hoses for damage e g kinks tears or leakage s contamination e g leaves hair dust or earth and deposits...

Page 11: ...the air from the dosing hose To do so connect pH FUN with the mains and start a manual dosing The purging is complete when you can no longer detect air bubbles on either the suction side canister pum...

Page 12: ...ecautionary measures need to be taken In particular it is very important that the entire system is protected against freezing temperatures and humidity Dosing system Rinse the pump with fresh lukewarm...

Page 13: ...ming a manual dosing The length of time or dose amount of the manual dosing can be set in the menu xx min Remaining time in minutes of a running manual dosing The remaining time is counted down during...

Page 14: ...provides an overview of all important current measurement and control parameters You can also set the operating mode in the normal view of the pH FUN Normally the pH FUN automatically ensures the opti...

Page 15: ...0 min Language German Language German Default Europe Initial operation Menu Normal operation 2 sec 2 sec If an alarm occurs a jump occurs automatically to the alarm display OK confirms all current pen...

Page 16: ...des Auto Man Man and Off are described in the table in section 4 1 Display and operating field of the pH FUN The execution of a manual dosing is additionally described in more detail in Chapter5 Perfo...

Page 17: ...The standard setting should only then be modified when the pH value of the pool water either does not reach the desired setpoint or overshoots the desired setpoint see also Chapter 9 3 The pH value i...

Page 18: ...does not activate as expected the time interval can be adapted accordingly see Chapter 9 1 The pH FUN reports an alarm Danger warning By setting the time interval to 0 minutes the dosing alarm is com...

Page 19: ...or Off You can abort a running manual dosing at any time by resetting the operating mode to Auto or Off 5 1 Calculating the manual dosing The following approximation formula can be used to calculate...

Page 20: ...the pH FUN device beforehand so that the electrode cable does not get twisted when unscrewing the electrode Handle the electrode by its plastic shaft only 5 Rinse the electrode in clean water and the...

Page 21: ...to play when calculating the dosing rate Parameter Explanation Current pH value in the pool water Desired pH setpoint The stronger the currently measured pH value of the pool water deviates from the s...

Page 22: ...m During a dosing alarm Only when the reason for the blocking is no longer present will the pH FUN again enable the dosing When dosing is blocked the pH FUN always shows a dosing rate of 0 in the norm...

Page 23: ...el of the canister in this case is no longer sufficient for an additional dosing the dosing is blocked Replace the empty canister The pH FUN subsequently resumes with the dosing Dosing alarm If the cu...

Page 24: ...f the pool water is too high An overdosing of pH Plus may possibly have occurred Reduce the value of the Min dos rate parameter see 4 3 3 Setting options in the menu in 3 steps to reduce the dosing ra...

Page 25: ...l overshoots the set setpoint An overdosing of pH Minus has occurred The dosing rate of the pH FUN is too high for the pool particularly in the area of the setpoint Reduce the value of the Min dos rat...

Page 26: ...correct measured values The pH7 buffer solution may possibly be contaminated and shows an incorrect pH value The correct sequence for the pH7 calibration may not have been adhered to precisely The pH...

Page 27: ...Level input signal 11 pH input signal 11 12 pH input to the connector of the pH electrode 12 pH input GND 10 2 Replacing the fuses There are 2 fuses on the circuit board No Fuse of Fuse F1 Power supp...

Page 28: ...ller and carefully remove the housing Be careful not to damage the circular seal inserted The fuses are located in a special holder and can be easily replaced The fuse holder is opened using a flat sc...

Reviews: