Instructions pH FUN V1
Page 16 of 28
4.3
Setting options of the pH FUN
This section describes all setting options offered by the pH FUN.
The setting of a parameter always proceeds according to the same principle. This is described as an example in
Chapter 4.1.2 Performing a parameter setting.
4.3.1 Menu navigation
To perform settings, please proceed as follows:
1. From the normal view, call up the menu by using a long press (at least 2 seconds) of the
button.
2. Initially, the menu page for the pH7 calibration appears (see 6 Performing the pH7 calib). Select the menu entry
that you would like to change by pressing the buttons
.
3. Using
, activate the input mode and perform the setting as described in 4.1.2 Performing a parameter setting.
4. After you have made all the desired changes, switch back to the normal view by using a long press (at least 2
seconds) of the
button again.
4.3.2 Switching the operating mode
You can switch between operating modes of the pH FUN directly from the normal view. By pressing the
button, the
input mode is activated and using the buttons
, the desired operating mode (Auto, Man.-, Man.+) is selected and
confirmed by pressing the
button again.
The following operating modes Auto, Man.-, Man.+ and Off are described in the table in section 4.1
Display and
operating field of the pH FUN
. The execution of a manual dosing is additionally described in more detail in Chapter 5
Performing a manual dosing.
4.3.3 Setting options in the menu
The menu tree of the pH FUN in section 4.2 The pH FUN menu already provides a complete overview of all setting
parameters in the menu of the pH FUN.
The following table provides an overview of all setting parameters in the menu, in each case with specification of the
allowed setting range, the setting step size as well as the default (standard) values for the maximum five different
default sets:
Default (standard) values for
default sets 1…5
Parameter
Setting range
Setting
step
size
Europe
UK
3
4
5
Manual dosing
0.1…4.0 litres
7…267 minutes
0.1 l
1,0
1,0
1,0
1,0
1,0
Dose amount
low
/
normal
/
high
-
normal
normal
normal
normal
normal
Min.dos.rate
0…25%
1%
5%
5%
5%
5%
5%
pH setpoint
6.0…8.0 pH
0.1 pH
7,20
7,40
7,20
7,20
7,20
pH alarm limits
Setpoint ± 0.1 pH
…
Setpoint ± 1.0 pH
± 0.1
pH
± 0.4
± 0.3
± 0.4
± 0.4
± 0.4
Acoust. alarm
On / off
-
on
on
on
on
on
Language
German / English /
…
-
D
D
D
D
D
Pool volume
1…10 m³
10…200 m³
1 m³
5 m³
40
40
40
40
40
Dos. direction
pH- / pH+
-
pH-
pH-
pH-
pH-
pH-
Dosing alarm
0…600 min
15 min
60
60
60
60
60
The default set can only be selected during the initial operation (see 4.4 Initial operation of the pH FUN). It cannot be
set in the menu.