background image

 

   

Instructions pH FUN V1

 

 

 

Page 11 of 28 

3.1.4  Replacing hoses 

To replace the hoses, please proceed as follows: 

 

Rinse the pump with fresh, lukewarm water for about 30 minutes. To do so, place the suction lances into a pail 
filled with tap water and start a manual dosing (see Chapter 5 Performing a manual dosing). 

 

Subsequently ensure that the system is fully disconnected from the power grid. This prevents the pump from 
switching on during the maintenance procedure. 

 

Remove the blue cover of the pump, loosen the sleeve nuts on the hose (2) and remove the suction and 
pressure hose. 

 

Press the band (1) together at the recessed grips and turn the catch (4) clockwise all the way. 

 

Move both hose ends outward and remove the band (1) with the hose (2). 

 

Pull the old hose (2) out of the two guides and insert the new hose.  

 

To mount the band (1) and hose (2), proceed in reverse order. Ensure that the hose (2) is firmly seated in the 
guides and that the catch (4) has engaged. 

 

As soon as all connections have been closed, you can purge the air from the dosing hose. To do so, connect 
pH FUN with the mains and start a manual dosing. 
The purging is complete when you can no longer detect air bubbles on either the suction side (canister – 
pump) or on the pressure side (pump – measurement cell for pH-Minus or injection piece for pH-Plus dosing). 

 

 

 

 

 

Figure: Head of the vacuum pump 

 

3.2 

Electrode information 

The electrodes must be replaced according to the maintenance plan annually or earlier if worn. Use original 
replacement electrodes only. These can be obtained from your swimming pool dealer under BAYROL Art. no.: 
185 301. 

Like the dosing hose, the pH electrode is also a wear part and is exempted from the guarantee/warranty. 

3.2.1  Electrode wear 

The following conditions, among others, indicate that the electrodes are worn: 

 

During the pH7 calibration procedure, the electrode takes unusually long to reach the value of the buffer 
solution. 

 

The deviation of the electrode during the pH7 calibration procedure is too large, the pH7 calibration is not 
possible. 

 

The white KCL salt rings in the electrode are used up or discoloured. 

 

    

 

ATTENTION:

    

Electrodes wear very quickly if there is an electrical potential on the pool water! 

Band 

Hose 

 

Pump body 

 

Catch 

 

 

Summary of Contents for pH FUN

Page 1: ...Instructions pH FUN V1 Page 1 of 28 Instruction manual pH FUN...

Page 2: ...ng field of the pH FUN 13 4 1 1 Button functions in the normal view 14 4 1 2 Performing a parameter setting 14 4 2 The pH FUN menu 14 4 3 Setting options of the pH FUN 16 4 3 1 Menu navigation 16 4 3...

Page 3: ...d operated by qualified expert personnel General danger warning Changing the system settings default values can be dangerous to persons under certain circumstances Therefore changes must only be made...

Page 4: ...ok after the disinfection The use of BAYROL algaecide and flocculant products is additionally recommended The product program is perfectly matched to guarantee a perfect water quality with very little...

Page 5: ...ck that the float in the measurement chamber can float freely Adjust the water flow through the cell so that the float just barely lies against the upper end in its guide bore Ensure that connections...

Page 6: ...lowed The equipment should be installed by a licensed electrician The supply voltage for the device may not exceed 240V 50 Hz The allowable operating temperature range is from 0 to 50 O C and the allo...

Page 7: ...to the measurement cell it must be fed directly into the circulating line You will require the following accessories available at your BAYROL dealer 100 509 PE hose length depending on the installatio...

Page 8: ...Measurement water intake 6 Heat exchanger 7 Sand filter 8 Measurement water extractor 9 Circulating pump 10 Container for pH Minus liquid 11 pH Minus suction line 1 pH FUN Chlorine 2 Swimming pool 3 M...

Page 9: ...available Use Europe for all countries not listed Confirm with Pool volume Enter the volume of the swimming pool in m cubic meter using the arrow keys and confirm with The input of the pool volume se...

Page 10: ...Performing the pH7 calib Replace the hoses of the dosing pumps 3 1 1 Checking the hoses Check all hoses for damage e g kinks tears or leakage s contamination e g leaves hair dust or earth and deposits...

Page 11: ...the air from the dosing hose To do so connect pH FUN with the mains and start a manual dosing The purging is complete when you can no longer detect air bubbles on either the suction side canister pum...

Page 12: ...ecautionary measures need to be taken In particular it is very important that the entire system is protected against freezing temperatures and humidity Dosing system Rinse the pump with fresh lukewarm...

Page 13: ...ming a manual dosing The length of time or dose amount of the manual dosing can be set in the menu xx min Remaining time in minutes of a running manual dosing The remaining time is counted down during...

Page 14: ...provides an overview of all important current measurement and control parameters You can also set the operating mode in the normal view of the pH FUN Normally the pH FUN automatically ensures the opti...

Page 15: ...0 min Language German Language German Default Europe Initial operation Menu Normal operation 2 sec 2 sec If an alarm occurs a jump occurs automatically to the alarm display OK confirms all current pen...

Page 16: ...des Auto Man Man and Off are described in the table in section 4 1 Display and operating field of the pH FUN The execution of a manual dosing is additionally described in more detail in Chapter5 Perfo...

Page 17: ...The standard setting should only then be modified when the pH value of the pool water either does not reach the desired setpoint or overshoots the desired setpoint see also Chapter 9 3 The pH value i...

Page 18: ...does not activate as expected the time interval can be adapted accordingly see Chapter 9 1 The pH FUN reports an alarm Danger warning By setting the time interval to 0 minutes the dosing alarm is com...

Page 19: ...or Off You can abort a running manual dosing at any time by resetting the operating mode to Auto or Off 5 1 Calculating the manual dosing The following approximation formula can be used to calculate...

Page 20: ...the pH FUN device beforehand so that the electrode cable does not get twisted when unscrewing the electrode Handle the electrode by its plastic shaft only 5 Rinse the electrode in clean water and the...

Page 21: ...to play when calculating the dosing rate Parameter Explanation Current pH value in the pool water Desired pH setpoint The stronger the currently measured pH value of the pool water deviates from the s...

Page 22: ...m During a dosing alarm Only when the reason for the blocking is no longer present will the pH FUN again enable the dosing When dosing is blocked the pH FUN always shows a dosing rate of 0 in the norm...

Page 23: ...el of the canister in this case is no longer sufficient for an additional dosing the dosing is blocked Replace the empty canister The pH FUN subsequently resumes with the dosing Dosing alarm If the cu...

Page 24: ...f the pool water is too high An overdosing of pH Plus may possibly have occurred Reduce the value of the Min dos rate parameter see 4 3 3 Setting options in the menu in 3 steps to reduce the dosing ra...

Page 25: ...l overshoots the set setpoint An overdosing of pH Minus has occurred The dosing rate of the pH FUN is too high for the pool particularly in the area of the setpoint Reduce the value of the Min dos rat...

Page 26: ...correct measured values The pH7 buffer solution may possibly be contaminated and shows an incorrect pH value The correct sequence for the pH7 calibration may not have been adhered to precisely The pH...

Page 27: ...Level input signal 11 pH input signal 11 12 pH input to the connector of the pH electrode 12 pH input GND 10 2 Replacing the fuses There are 2 fuses on the circuit board No Fuse of Fuse F1 Power supp...

Page 28: ...ller and carefully remove the housing Be careful not to damage the circular seal inserted The fuses are located in a special holder and can be easily replaced The fuse holder is opened using a flat sc...

Reviews: